Alpha 4314 Series Standard (STD) and Reduced Current (KRC) Models Digital Igniter Tester Operations Manual Valhalla Scientific, Inc. | 12127 Kirkham Rd. Poway CA 92064 | www.valhallascientific.com...
Page 2
The warranty period for this instrument is stated on your invoice and packing list. Please refer to these to determine appropriate warranty dates. We will repair the instrument during the warranty period provided it is returned to Valhalla Scientific, Inc. freight prepaid. No other warranty is expressed or implied.
Page 3
REMOVE POWER and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to Valhalla Scientific for service and repair to ensure that safety features are maintained. DO NOT service or adjust alone: Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Instrument Identification Valhalla Scientific instruments are identified by a two part serial number. The Serial Tag is located on the rear or bottom of the instrument. The number is in a form of 00-0000. The first two digits, called the serial number prefix, indicate the model.
If the unit fails to operate or fails to meet the performance specifications of Chapter 3, notify the carrier's agent and the nearest Valhalla Sales Office. Retain the shipping carton for the carrier's inspection. DO NOT return equipment to Valhalla Scientific or any of its sales offices prior to obtaining authorization to do so.
Accuracy Table 3 Range Accuracy 2Ω ±0.02% of reading ± 0.02% of range 20Ω ±0.02% of reading ± 0.02% of range 200Ω ±0.02% of reading ± 0.02% of range 2kΩ ±0.02% of reading ± 0.02% of range 20kΩ ±0.02% of reading ± 0.02% of range 200kΩ...
Chapter 4 – Optional Equipment Available Options Listed below are the options available for use with the Alpha 4314 Series Digital Igniter Testers. Option A: Battery Charger Option "A" is an AC/DC converter that converts 115VAC line voltage to 6VDC at 300mA. One charger is provided as a standard accessory with every Alpha 4314.
Option MP-3: Kelvin Micro-Probes Option "MP-1" is a 48" shielded lead set (dual banana), handheld pencil type, terminated in spring loaded steel tips with .08" separation (1 amp max.). Special orders or modified versions available upon request. Option MP-4/MP-5: Surface Probes These probes permit rapid, repeatable bonding testing on a variety of screened or flat surfaces.
Chapter 5 - Operations General This section of the manual contains complete operating instructions for the Alpha 4314 Series Digital Igniter Testers. A description of the front panel controls, connection instructions, and the theory behind 4-wire resistance measurement is discussed in this section.
4-Wire Resistance Measurement The four-terminal configuration of the Alpha 4314 eliminates errors normally caused by test lead and contact resistances. In many applications the contact resistance can exceed the value of the load by several orders of magnitude. The Alpha 4314 bypasses this potential error source by providing two terminals of constant current and an additional two terminals for high impedance voltage measurement.
LED marker on the Alpha 4314 front panel. Every Alpha 4314 Series Igniter Tester is thoroughly tested before it leaves the factory. These tests include a failsafe test that simulates a "worst case" failure condition. The resulting output current is recorded and provided on the Certificate of Calibration included with every Alpha 4314.
Battery Monitoring Circuitry A standard feature of Alpha 4314 is a circuit that monitors the output voltage of the internal battery supply. The Battery Status LED for this feature is located to the far left of the display window. • A green Battery Status LED indicate that the batteries are charged and the device can be used.
Use only the specified component or its exact equivalent. Spare parts can be ordered from your nearest Valhalla Scientific Service Center or directly from the factory by referring to the Valhalla Stock Number.
Chapter 7 - Routine Maintenance General This section of the manual contains routine maintenance information regarding the Valhalla Scientific Model Alpha 4314 Digital Igniter Testers. Calibration should be performed regularly to ensure continued instrument accuracy. The recommended calibration interval is one year.
Calibration Procedure VOLTAGE SENSE ADJUSTMENTS for STANDARD CURRENT MODELS Connection: Apply voltage at Voltage sense Input and short current source and Voltage sense negative Terminals. And select second Range 1. Apply to the Voltage Sense 0.000mV using a precision power supply. 2.
RESISTANCE ADJUSTMENTS 1. Select 2 range. 2. Connect a standard resistor ½ scale of the selected range. Example: If the 200Ω range is selected connect to a 100Ω standard resistor. 3. Adjust R15 for a display reading of “10000” (Decimal point unimportant). 4.
The rechargeable NiMH batteries used in the Alpha 4314 are very durable and should provide years of trouble-free operation. As with all batteries, a replacement will eventually be necessary. Batteries may be ordered from Valhalla Scientific as stock #05-10117, quantity: (4). The process of battery replacement is described below: Remove the four feet screws and the bottom cover.
Chapter 8 - Interface Operations BCD Interface Option BCD has a rear mounted 50 pin Amphenol connector providing data in a parallel binary-coded- decimal (BCD) format. The outputs are TTL compatible and may drive 1 LS load. The summary of pin functions is provided below and on the schematic number 2053-076 at the back of this manual.
Idn? The device should return the Manufacturer, model number, firmware version in the following format: Valhalla Scientific Alpha4314 V2.0 If you do not receive a proper response from the Alpha 4314, please check if the power is on, and all...
LogOff Reset LogON Ohm? Screen? LogOn0 Ohms? LogOn1 RangeN? Idn? QUERY Returns the RS-232 identification string from non-volatile. Syntax: Idn?<crlf> Response: ID string "Valhalla Scientific Alpha4314 v2.0" Example: Idn? "Valhalla Scientific Alpha4314 v2.0" ID_STRING "Valhalla Scientific" MODEL "Alpha4314" VERSION "V2.0"...
Page 25
Local COMMAND Returns meter to local mode. Syntax: Local<crlf> Returns: <crlf> Power-on default = LOCAL mode Notes: REMOTE mode is selected when the meter receives a valid character (not <crlf>. LogOn COMMAND Turns on continues reading mode synonymous with “Ohms?”. Note: if the devices is already Logging, the LogOn command will return “Invalid Command”...
Page 26
LogOn1 COMMAND Turns on continues reading mode synonymous with “Ohm?”. Note: if the devices is already Logging, the LogOn1 command will return “Invalid Command” Syntax: LogOn1<crlf> Returns: Measurement Value expressed in Ohms every 500ms. Example for 2Ω Range: 1.2345 Example for 20Ω Range: 12.345 Example for 200Ω...
Page 27
Ohm? QUERY Responds with the front panel reading formatted in ohms. No leading zeros. Syntax: Ohm?<crlf> Returns: Display data in Ohms. Example for a 1.1 ohm reading on: 2Ω Range: 1.1000 20Ω Range: 1.100 200Ω Range: 1.10 2kΩ Range: Ohms? QUERY Responds with the front panel reading formatted in ohms with multiplier.