Local
Returns meter to local mode.
Syntax:
Local<crlf>
Returns:
<crlf>
Power-on default = LOCAL mode
Notes: REMOTE mode is selected when the meter receives a valid character
(not <crlf>.
LogOn
Turns on continuous reading mode synonymous with "Ohms?". Note: if the
devices is already Logging, the LogOn command will return "Invalid
Command"
Syntax:
LogOn<crlf>
Returns:
Measurement Value + Multiplier every 300ms.
Example for 2Ω Range:
Example for 20Ω Range:
Example for 200Ω Range:
Example for 2kΩ Range:
Example for 20kΩ Range:
Example for 200kΩ Range:
Example for 2MΩ Range:
Example for 20MΩ Range:
LogOn0
Turns on continuous reading mode synonymous with "Ohms?". Note: if the
devices is already Logging, the LogOn0 command will return "Invalid
Command"
Syntax:
LogOn0<crlf>
Returns:
Measurement Value + Multiplier every 500ms.
Example for 2Ω Range:
Example for 20Ω Range:
Example for 200Ω Range:
Example for 2kΩ Range:
Example for 20kΩ Range:
Example for 200kΩ Range:
Example for 2MΩ Range:
Example for 20MΩ Range:
COMMAND
COMMAND
COMMAND
1.2345
12.345
123.45
1.2345k
12.345k
123.45k
1.2345M
12.345M
1.2345
12.345
123.45
1.2345k
12.345k
123.45k
1.2345M
12.345M
24