The user can easily detect the mode selected by looking at the symbol in the top right corner of the display.
Indicates that the device is in Battery Mode
Indicates that the device is in Continuous Power Mode.
When in Battery Mode, the 4314Ki run of 4 NiMH D Cell batteries. The batteries are charged only while
the device is in battery mode and turned off. Internal safety line isolation switch disables the battery
charging circuit when the 4314Ki is ON. All power is removed from the source circuitry when the unit is
OFF and in charging mode.
In Continuous Power Mode, the 4314Ki is powered via the provided power supply (also used as a battery
charger). This configuration is a safe option for applications that require longer testing times.