ELECTROMAGNETIC COMPATIBILITY (EMC)

• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
• Recommended separation distances between portable and mobile RF communication equipment
and the device.
The device is intended for use in an electromagnetic environment where radiated RF disturbances are
under control. User can help prevent electromagnetic interference by keeping the device at a minimum
distance from portable and mobile RF communications equipment (transmitters). Below table details
the maximum output power of transmitter:
Rated maximum
output power of
IB-WGNBPW-710_A--01.indd 26
Compliance
Group 1
Class B
Not
Applicable
Not
Applicable
IEC 60601 test level
± 6 kV contact
± 8 kV air
3 A/m
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
transmitter W
Not Applicable
0.01
NA
0.1
NA
1
NA
10
NA
100
NA
Electromagnetic environment – guidance
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not
likely to cause any interference in nearby electronic
equipment.
The device is suitable for use in all establishments,
including domestic establishments, and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Compliance level
Electromagnetic environment–
guidance
± 6 kV contact
Floors should be wood, concrete
± 8 kV air
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
26
800 MHz to 2.5 GHz
d = 2.3
0.23
0.73
2.3
7.3
23