Emissions
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Premium Plus is intended for use in the electromagnetic
environment listed below. The user of the device should ensure that it is used in such an environment to prevent
improper operation.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Rated power
frequencymagnetic fields
IEC 61000-4-8
Electrical fast transient/ burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Conducted disturbances
induced by RF fields
IEC 61000-4-6
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Compliance
Class B
Complies
Class A Category
Complies
Immunity test level
±8 kV contact
±15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
385 (27), 450 (28), 710 (9), 745
(9), 780 (9), 810 (28), 870 (28),
930 (28), 1720 (28), 1845 (28),
1970 (28), 2450 (28), 5240 (9),
5500 (9), 5785 (9) MHz (V/m)
30 A/m
50 Hz or 60 Hz
±2 kV
100 kHz repetition frequency
±1 kV Line-to-line
±2 kV Line-to-ground
3 Vrms, before modulation is
applied
0.15 MHz – 80 MHz
6 Vrms, before modulation is
applied, in ISM and amateur
radio bands between 0.15 MHz
and 80 MHz
80 % AM at 1 kHz
0 % UT; 0.5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°and 315°
0 % UT; 1 cycle and 70 % UT;
25/30 cycles single phase at 0°
0 % UT; 250/300 cycle
Electromagnetic environment – guidance
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power supply
network which supplies buildings used for domestic
purposes.
Electromagnetic environment - guidance
Floors can be a source of static electricity. When the air
is very dry (<20 % RH) the user is advised to prefer to
use the device in a room with concrete or wooden floors
(instead of using it in a room with a synthetic floor).
The device is immune against any RF wireless
communication device such as smart phone, Wi-Fi
or Bluetooth device at a distance of 30 cm or more.
Avoid a closer distance as this may result in improper
operation.
Mains power quality expected of a typical commercial
or residential environment.
Mains power quality expected of a typical commercial
or residential environment.
Mains power quality expected of a typical commercial
or residential environment. At power interruptions the
breast pump will automatically switch-over to battery
mode without interruption of the function.