Installation
3
2
3
Fig. 4

Installation of the printer in the monitor trolley

NOTICE
Pull out the two guide rails completely to the front.
Place the printer on the two guide rails.
Fig. 4 shows the two drawn-out guide rails with the attached printer from below.
Place the power chain (1/Fig. 4) in a semicircle around the angle bracket (2/Fig. 4).
Secure the printer to the rails with the 12 screws supplied.
The 3 rear threads on each of the rails can be accessed by pushing the printer through
the openings (3/Fig. 4).
Push the printer all the way into the monitor trolley and at the same time guide the power
chain through to the rear.

Attaching the power chain

Attach the bracket (2/Fig. 5) at the printer with two screws.
Unscrew the two screws (1/Fig. 5) from the angle bracket (2/Fig. 5).
Attach the power chain (3/Fig. 5) to the bracket (2/Fig. 5) with the two screws (1/Fig. 5).
Siemens AG
Medical Solutions
1
One person should hold the printer to make sure it does not fall,
while a second person tightens the screws.
SPR2-130.814.01
Rev.01_V 03.01
1
Fig. 5
Page 3 of 14
TD PS 24
2 - 3
2
3
SIREMOBIL Compact
System Manual
2
2

Advertisement

Table of Contents
loading