Installation When Upgrading In The Field; Replacing Fuse F5; Checking Or Replacing Fuse F5 - Siemens SIREMOBIL Compact CODONICS EP 1660 System Manual

Table of Contents

Advertisement

2 - 6

Installation when upgrading in the field

WARNING

Replacing fuse F5

Requirements
The system must be isolated from the power supply. Unplug the power connector from
the power socket.
Open the rear cover of the monitor trolley.

Checking or replacing fuse F5

Check the value of fuse F5 on board D50 of the ON/OFF assembly (installed in the
bottom shelf of the monitor trolley).
If a 4 AT fuse is fitted, replace it with the 6.25 AT fuse supplied.
On board D50, stick the label (6.25 AT) supplied on top of the fuse label (4 AT).
Manually change the fuse value of the F5 fuse to 6.25 AT in the wiring diagram.
Enter the remark "Condonics printer installed" together with the date and your initials.
SIREMOBIL Compact
System Manual
The system must be isolated from the power supply before begin-
ning work. Unplug the power connector for the system from the
power socket!
SPR2-130.814.01
Rev. 01
Page 6 of 14
03.01
TD PS 24
Installation
2
2
2
Siemens AG
Medical Solutions