Page 1
Overview The MIQ/IC2 current input module provides two 0/4 - 20 mA current inputs for IQ SensorNet and thus enables to connect external sensors via their current output. Measured values of the external sensors can be displayed, recorded and processed like the measured values from IQ SensorNet.
Page 2
Warning This document is not intended to replace the MIQ/IC2 module operation manual. Please use the operating manual as a reference during the following functions; installation, operation, cleaning, maintenance and troubleshooting. Step By Step Instructions Install all cable glands to the bottom of the MIQ/IC2 module as shown below Connect the cable to the terminal strip Open the MIQ/IC2 module.
Page 3
Loop Powered Devices: The 24V power supply in the MIQ/IC2 is rated for 240mA. If powering two devices make sure the current draw of the entire 24V loop (the devices and isolators) Figure 1: Press the “S” button on the IQ SensorNet controller doesn’t exceed 240mA.
Page 4
Scroll down to the “Setting of sensors and Select the “Measuring range” column and diff. sensors” option and press “OK”. press “OK”. Press “Continue” Customize the following settings to the appropriate values...
Page 5
Measuring range Decimal places Display value (0/4mA) Display value (20 mA)
Page 6
Display unit Measured parameter Error detection Error threshold...