3-6
Operation
Configuring scan parameters
This paragraph details how to select internal scanning and how to configure the scan
parameters.
Front Panel
Use the SHIFT-CONFIG keys to select the scan type and configure the scan parameters.
1.
2.
3.
4.
5.
Press SHIFT and then CONFIG. The instrument will display the following prompt for
internal scanning:
TYPE: I20
Use the L or M key if EXT is displayed and press ENTER.
The following message will display:
MIN CHAN: 01
Use the L and M keys to select the channel at which the scanning will begin and press
ENTER.
The unit will display the following message:
MAX CHAN: 20
, L, and M keys to select the channel at which the scanning will stop
Use the
,
and press ENTER.
The instrument will display the following message:
TIMER? OFF
Use the
or
key to select the ON option if desired and press ENTER. Select the
scan time when prompted and press ENTER:
00H:00M:01.000S
The unit will display the following message:
RDG CNT: 0020
, L, and M keys to select the desired number of readings from the
Use the
,
selected list to be taken and stored by the instrument and press ENTER.
The instrument will store in memory the number of readings specified with RDG CNT.
After the scanning has been completed, you can access these readings by pressing the
RECALL key on the front panel and using the
the stored readings.
, L, and M keys to scroll through
,