Opening channels

Front panel
Use the front panel OPEN key to open all the card channels.
Remote interface
Use the :ROUTe:OPEN:ALL command to open all the channels. The command is as follows:
PRINT #1, "output 16; :route:open:all"
Use the :ROUTe:MULTiple:OPEN command to open specified channels. For example, if you
want to open channels 1, 3, and 5, use the following command:
PRINT #1, "output 16; :route:multiple:open (@ 1,3,5)"
NOTE
Opening channel 21 selects the 2-pole operating mode.

Selecting measurement function

Use the following procedure to select a specific measurement function for an internal
channel:
1.
Use the
2.
Select the desired measurement function with the keys on the front panel. The various
measurement functions available are:
DCV: DC volts
ACV: AC volts
Ω2W: 2-wire ohms
Ω4W: 4-wire ohms
(channels 1 through 10)
FREQ: frequency
and
keys to select the desired channel.
PERIOD: period
TEMP: simulated temperature
CONT: continuity
: diode test