Typical Connecting Schemes; Voltage Connections - Keithley 2000-SCAN-20 Instruction Manual

Scanner card
Table of Contents

Advertisement

2-8
Card Connections and Installation

Typical connecting schemes

The following paragraphs discuss typical connections for the scanner card.

Voltage connections

Figure 2-4
used in the 2-pole mode and that up to 20 voltage sources can be switched with this
configuration. This basic configuration can be used for the following types of measurements:
All channels (1 through 20) can be used with this configuration.
Figure 2-4
Connections for voltage scanning
2000-SCAN-20 Card
shows typical connections for voltage measurements. Note that all channels are
DCV
ACV
Frequency (voltage only)
J1036
HI
CH 1
LO
HI
CH 20
LO
J1035
HI
OUT A
LO
HI
OUT B
LO
HI
V
LO
Voltages
Under Test
HI
V
LO
HI
To DMM
LO
Input