In the US, federal law restricts this system to sale by or on the order of a physician. Indications for use The Aria Free system by Aria Health is intended for use by medical professionals and patients at home for the treatment of the following conditions: •...
Contraindications Use of pneumatic compression may be contraindicated in some patients with the following pre-existing conditions: • Congestive heart failure • Any condition where increased lymphatic and venous return is undesirable Adverse effects The following side effects may arise during the course of using the system: •...
Page 4
Your Aria Free system at a glance The Aria Free system includes the following: A. Air pump B. Pneumatic garment C. Connector tube D. Power supply unit...
Page 5
Your Aria Free system in detail Air pump Start/Stop button button Bluetooth ® Air pump connection point Air filter cover Pneumatic garment Garment connection point Self-fastening garment tabs Connector tube Air line Power connection point Electrical line Power supply unit Wall plug Power line English...
Page 6
Your assembled system Your fully assembled system will look similar to below. Detailed Setting up and assembly instructions can be found in the section starting therapy. Aria app Your Aria Free system is designed to work with the Aria app. However, the Aria app is not required to operate the system.
Setting up and starting therapy If the app has been downloaded, open it and follow the step-by-step instructions. You will also find helpful videos to assist you in your first time setup. Use your Aria Free system only as directed by your physician or healthcare provider.
Page 8
2. Plug the power line of the power supply unit into the remaining port on the grey power connection point. Plug the other end of the power supply into the power outlet on your wall. 3. After following these steps, your air pump connections should look like this.
Page 9
Wearing your garment WARNING Do not put on the garment if your affected leg has open wounds. The system is only intended to contact intact skin. Ensure any wounds are adequately covered with wound dressing prior to use. Prepare the garment by loosening the self-fastening tabs from ankle to thigh but do not remove the tabs around the foot sole.
Page 10
Your garment should have a tight but not restrictive fit, like a yoga pant. It is important to avoid wearing anything underneath your garment that may hamper the lymph flow. These items include: pants with heavy seams, compression stockings, zippers. You may wear loose-fitting, non-elastic clothes, sheer liners or place the garment directly on intact skin.
Page 11
Connect the garment 1. Connect the green end of the airline and the electrical line to the garment connection point as shown on the right. 2. The garment connection point should look as shown on the right. Using therapy Once all the connections have been made, you are ready for therapy Start therapy by pressing the Start/Stop button.
Page 12
Therapy session One full therapy session will take sixty (60) minutes. Once you press the Start/Stop button, therapy will commence. You will feel the garment inflate starting from the foot, and gradually progressing up to the thigh. Once the garment is fully inflated, it will gradually deflate starting from the thigh, and continuing down to the foot.
Ending therapy Once your full therapy session is complete, the air pump will stop inflating and a green light will appear on the indicator light above the power button. You can then follow the below steps: Remove the garment from your leg. Leave the garment with the tube and air pump connected for at least 5 minutes.
Page 14
Removing air from your garment A deflation mode is available in your Aria Free system. It is designed to allow you to remove trapped air from the garment. Follow the below steps to enter this mode and use it. 1. Connect your device as you would for therapy. However leave the garment tabs unfastened and disconnect the green end of the air line from the garment.
Page 15
Using your system with the Aria app Your Aria Free system can be used with or without the Aria app by Aria Health. The Aria app will give you additional data about your therapy sessions. When you open the app for the first time, you will see instructions on how to pair the app with your system.
Page 16
5. Consent to the privacy policy and terms of use and to allow the Aria app to upload my data is required to use the app. • To accept Aria Health terms of use and privacy notice, select that check box.
Page 17
8. You will be taken through a series of short videos that will describe how to connect and power up your Aria Free system, as well as properly fit your garment. If you want to revisit these Help Bar setup videos later, they are available on the page under ‘Aria Setup’.
Page 18
Done When successfully connected, the light on your device will be Therapy a solid blue and it will take you to the page. Bluetooth status indicators Bluetooth indicator light is flashing blue. Your machine is in discoverable mode and ready for pairing. Bluetooth indicator light is constantly Your machine is connected to your blue.
The air pump has an expected life of five years based on normal daily usage. Please contact Aria Health if any deterioration in air pump performance (e.g. failure to inflate air chambers) is observed.
Page 20
Remove the garment 1. Loosen the self-fastening straps. 2. Remove your limb from the garment. 3. Flatten the garment for cleaning by opening all of the self- fastening straps and placing the garment on a flat surface (with the side you want to clean facing up). WARNING •...
Page 21
CAUTION • If any visible deterioration of a system component is apparent (cracking, discoloration, tears, breaks in tubing etc.), the component should be discarded and replaced. • If the leg garment is damaged (rips, tears in fabric or tubing) or has become excessively soiled and the soiling cannot be cleaned, it is recommended that the garment be discarded.
Page 22
Cleaning the garment • Wipe clean the interior of the garment with a damp cloth each week. • Using a clean cloth, lightly moisten it with warm water to dampen. Mild detergent may be used to address difficult soiling. • Wipe clean the interior surface of the garment using the lightly dampened cloth.
Page 23
Change the Air pump filter Replace the filter at least every 6 months. It cannot be washed. Check and replace it more often if there are any holes or blockages caused by dirt or dust. To change the filter: 1. Slide the filter cover off and remove the old filter. 2.
Page 24
Traveling You can take your Aria Free system with you wherever you go. It has a universal power supply that will accept voltage standards across the world. However, make sure you have the appropriate power plug adapter for the region where you will be traveling. Please contact us or visit our website for official Aria travel accessories.
Troubleshooting If you have any problems, consult the following troubleshooting topics. If you are not able to fix the problem, contact Aria customer support. Do not try to open the machine. Problem/possible cause Solution Air pressure in my garment seems too low / It feels like I am not getting enough air.
Page 26
Inspect the connector tube for any damage (cracks, tears, breaks). If the tubing is damaged, do not attempt to repair it. Please contact Aria Health for a replacement garment. Inspect the leg garment for damage (rips, There may be damage to your garment tears in fabric or tubing).
• Do not open or modify the system. There are no user-serviceable parts inside. Repairs and servicing should only be performed by an authorized Aria Health service agent. • Beware of electrocution. Do not immerse the air pump, garment, power supply, or power cord in water. If liquids are spilled into or onto the system, unplug the system and let the parts dry.
Page 28
CAUTION • Use only Aria Health parts and accessories with the system. Non- Aria Health parts may reduce the effectiveness of the treatment and/or damage the machine. • Blocking the air tubing and/or air inlet of the machine while in operation could lead to overheating of the machine.
Technical specifications System Pressure range Normal: 0-45mmHG Single Fault: <70mmHG Materials Flame-retardant engineering thermoplastic Operating temperature +41°F to +95°F (+5°C to +35°C) Note: Under extreme ambient temperature conditions (104°F/40°F) the tubing near the garment may reach temperatures up to 109°F/43°C. These parts of the tubing are not in contact with the user and machine remains safe under these extreme conditions.
Page 30
Additional information regarding the FCC Rules and IC compliance for this machine can be found on www.Aria-therapy.com. Aircraft use Aria Health confirms that machine meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel. Wireless technology...
Page 31
Design life 5 years (controller/pump, based on daily usage of up to 2 hours per day) Based on continued cleanliness (compressible limb sleeve garment) General The patient is an intended operator. Notes: • The manufacturer reserves the right to change these specifications without notice.
Page 32
Bluetooth (Bluetooth Brand guide) Prescription only (In the US, Federal law restricts these devices to sale by or on the order of a physician) (21 CFR 801) Operating altitude Atmospheric pressure limitation (ISO 15223-1) Complies with RTCA DO-160 section 21, category M. MR unsafe (do not use in the vicinity of an MRI device) (ASTM F2503) Replace 30 days after opening.
If you have a medical question, please contact your healthcare provider. Limited warranty Aria Health warrants that your Aria Health product shall be free from defects in material and workmanship from the date of purchase for the period specified in the following table:...
Page 34
If the product fails under conditions of normal use, Aria Health will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product;...