Omron HBP-1320 Instruction Manual page 11

Hide thumbs Also See for HBP-1320:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
2829490-0B_IM-HBP-1320-E-03-04-2020.book Page 9 Thursday, April 1, 2021 8:24 AM
Note:
Setup
• Read and understand the manual for each optional accessory. This manual does not contain
cautionary information for optional accessories.
• Exercise caution with the cables and arrange so that the patient does not become entangled or
bound.
Before use / during use
• Check the following after turning on the power:
- No smoke, abnormal odor, or abnormal noise is emitted.
- Press each button and verify that it operates.
- For functions that cause icons to light or flash, verify that the icons light or flash (page 13).
- Measurement can be performed normally, and measurement error is within the tolerance value.
• If the screen is not displayed normally, do not use the unit.
• When recycling or disposing of parts (including batteries) of the unit, follow local government rules
and regulations.
Cleaning
• For cleaning, see page 24.
Rechargeable battery
• To prevent accidents, keep batteries out of reach of infants and small children.
• If you sense that something is wrong with a battery, immediately move it to a safe location and contact
the administrator responsible for the unit or your OMRON retail dealer or distributor.
• If the battery voltage is low, operation by battery may not be possible.
Non-Invasive Blood Pressure (NIBP) measurement
• If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where
the cuff is to be wrapped, follow the instructions of the doctor.
• Non-Invasive Blood Pressure measurement (NIBP) is performed by compressing the upper arm.
Some people may experience intense pain, or transient spotting caused by subcutaneous
hemorrhaging (bruising) may appear. The spotting will disappear with time; however, it may be
appropriate to inform patients for whom this may be a concern that spotting sometimes occurs, and if
necessary, refrain from measurement.
• Clinical evaluation of the unit was carried out in a resting position (feet on floor without crossing legs,
back supported by backrest, arm supported such that cuff is attached at same height as heart) in
accordance with ISO81060-2:2013. Measurement in a position other than the resting position may
result in differences in measurement results.
• As the patient's posture or physiological condition influences blood pressure, it easily fluctuates.
Recommend the followings to take a more accurate measurement:
- Make the patient relaxed.
- Do not talk to the patient.
- Let the patient rest for at least 5 minutes before the measurement.
EN
9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents