Step 1: Insert probe intended to be used with ultrasound guidance. The BD EleVation™ Probe may be inserted into the driver in either of the following ways: TriConcave™ Tip OPTION 1: Remove the BD EleVation™ probe from the package using aseptic technique by grasping the BD EleVation™...
Ensure the sample container is secure before acquiring additional samples. Cleaning and maintenance • After each use, wipe the BD EleVation™ Driver with a Super Sani-Cloth or a cloth dampened with water to remove any excess blood or fluids present on the device.
Page 4
4. Place the BD EleVation™ Driver in the wireless charging stand. The BATTERY indicator on the BD EleVation™ Driver will illuminate per the table to indicate the battery is charging.
• Remove BD EleVation™ Probe Error during: ERROR indicator blinks • Place BD EleVation™ Driver on a flat • Calibration red, PRIME/PIERCE and surface with hand away from any SAMPLE buttons are • PRIME/PIERCE...
Page 6
Biopsy System should not be used in an operating room. 6. The BD EleVation™ Breast Biopsy System is not classified as an AP or APG device. 7. The BD EleVation™ Breast Biopsy System is not suitable for use in the presence of flammable anesthetic. 8. The BD EleVation™ Breast Biopsy System is not suitable for use in an oxygen rich environment.