Page 1
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 - US Operator Manual...
Page 2
All manuals and user guides at all-guides.com...
Page 3
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Copyright Note The information contained in this document is the property of VWS (UK) Ltd. and is supplied without liability for errors or omissions. No part of this document may be reproduced or used except as authorized by contract or other written permission from VWS Ltd.
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA TABLE OF CONTENTS Use of this Manual ............3 Customer Support ............3 HEALTH AND SAFETY NOTES ......... 3 Electricity ..............3 Pressure ..............3 Sanitization Chemicals ..........4 Control of Substances Hazardous to Health (COSHH)4 Consumables.
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual INTRODUCTION Use of this Manual ELGA LabWater strive to produce manuals that are as simple and accurate as possible. However, should you feel they can be improved in any way please email us at info@elgalabwater.com This manual guides you through the installation, commissioning and...
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Sanitization Chemicals During the sanitization cycle Minncare Cold Sterilant is used and relevant safety guidance is included in this handbook. Please refer to the manufacturer for material safety data sheets. Minncare Cold Sterilant is EPA registered as a sterilant, high level disinfectant, and sanitizer.
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual INSTALLATION AND COMMISSIONING Installation The unit is supplied with a quick start manual that allows the unit to be quickly installed and commissioned. It is recommended that a trained ELGA representative carry out the installation of the product.
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA CONTROLS The MEDICA control panel has a range of control icons. General icons are as follows. Further icons are described in the appropriate sections a complete listing is included in Section 11. BUTTON ICON Function...
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual User Recognition Keys Your unit is supplied with the following User keys: Master Key (Black) 1off User Key (Blue) 4off Sanitization Key (Green) 2off The Master Key should be stored in a safe place. The Master Key controls the access level of the other User Keys.
Page 10
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Step 2 - Clock (screen 018) PRESS to edit time PRESS ✔ to proceed to step 3. PRESS and HOLD to increase or to decrease hour.
Page 11
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Step 5 –Set units of measure - Volume (screen 021) This screen allows units of water volume to be set, to either Litres or US Gallons. This only indicates the volume in the reservoir.
Page 12
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Step 9 - RO water quality alarm settings QS1 (screen 025) This screen is used to set the value at which the RO water quality alarm will sound. This alarm does not stop RO production but can extend or start flush routines.
Page 13
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Step 12 - Product water temperature alarm settings TS2 (screen 028) This screen is used to select the value at which the product water temperature alarm will activate. The alarm does not stop the unit and will automatically reset if the temperature returns below the set point.
Page 14
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA PRESS to increase or to decrease time in increments of 30 minutes. PRESS to enter times. PRESS ✔ . Note: Night service will only be operational if Continuous Operation is not enabled.
Page 15
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Step 17 - Auto-restart (screen 033) This allows selection of the AUTO restart option. If auto restart is selected the unit will automatically restart after a power failure.
Page 16
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Step 21 - Programming of User Keys (screen 037) During the life of the MEDICA you may need to delete or add User keys to prevent or allow access to user choices. This feature is only available to the Master key holder.
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Setting Consumable Replacement Reminders Step 1 - Enter consumable replacement timer set-up TURN unit off at electrical supply. Note: Allow time for the display to go blank after the power is removed.
Page 18
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Step 5 - Protek L1/L2 replacement date PRESS button to reset Protek L1/L2 replacement date PRESS ✔ to accept replacement date and proceed to UMF replacement. PRESS ✔...
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual OPERATION Night service/Operational day (screen 030) Refer to Section 3.2 - Step 14 and 15. Your MEDICA can be programmed to operate on specific days between selected times. This is to optimize the efficiency of the unit and to minimize rises in water temperature.
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA MAINTENANCE An approved supplier or distributor should carry out any maintenance work not included in this handbook. Note: Disposal of all end of life consumable items should be in accordance with local statutory regulations WARNING! ALWAYS CHECK...
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual SWITCH on power. FOLLOW screen prompts to accept new cartridge replacement date START unit and allow to circulate until water quality is achieved. CAUTION! If operating the unit with two Medpure L1 purification packs installed, both packs must be replaced at the same time.
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Replacing UV lamp (LC105) The UV lamp should be replaced in the following circumstances: When indicated by the consumable alarm. After a maximum of twelve months use. ENSURE process is OFF and ISOLATE power.
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Protek L1/L2 Replacing Protek L1 (LC175) or Protek L2 (LC177) Pre-treatment Cartridges The replacement frequency of the pre-treatment cartridge is dictated by the purity of the feed water. It should be replaced in the following circumstances: ...
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA Replacing E- Cartridge (LC181) standard on RE variant only The E- cartridge should be replaced: When indicated by the alarm E- Cartridge 1. ENSURE process is OFF and ISOLATE power. 2.
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual SANITIZATION RO sanitization (Step 1) The RO is sanitized to reduce the growth of microbiological contamination within the RO module. The MEDICA has a built-in reservoir which allows the sterilant to be introduced into the water feeding the RO.
Page 26
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA PRESS ✓ to enter RO sanitization. Enter RO sanitization UNSCREW the large white lid on the left of the unit. ADD 50ml of Minncare Cold Sterilant REPLACE lid.
Page 27
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual The system now enters a period of semi automatic sanitization lasting approximately 90 minutes ALLOW the first timer to progress down until it displays just 5 minutes. PRESS the process button and the unit should pause.
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA EMERGENCY BYPASS Contact your local service provider before putting the unit into emergency bypass. Emergency bypass (MEDICA R/RE/RX) Ensure the process is OFF Disconnect the electrical supply Open the doors.
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Calculation for Medpure L1 Cartridge life Calculate the maximum Medpure cartridge life under emergency bypass processing and with your mains water conditions to maintain a supply purity of 1MΩ.cm to the analyzer The maximum life expectancy of the Medpure L1 cartridge in emergency bypass mode while maintaining a supply purity of 1MΩ...
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA TROUBLE SHOOTING If a problem occurs the unit will normally sound an alarm and the respective icon will flash. The audible alarm can be silenced by pressing the mute button. The appropriate action to then take (for the more common alarm displays) is shown on the Quick Reference Guide (fitted to the inside of the door of the MEDICA.) If the unit cannot be repaired using the information shown, please call your local ELGA LabWater...
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual KEY TO CONTROL PANEL ICON DESCRIPTION ICON DESCRIPTION User Accept Recognition Auto restart Locked Scroll back Menu Step back Mute alarm Bell Night Standby Option OFF Calibration Option ON point...
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA TECHNICAL SPECIFICATION Feedwater Source Potable tap water as detailed below. Failure to comply with the minimum feedwater pretreatment recommendations will affect the life and performance of key components within the MEDICA, and may invalidate the warranty.
Page 33
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual Connections Inlet 10 mm OD tube Drain 10 mm OD tube Reservoir drain 10 mm OD tube Recirculation loop outlet* 10 mm OD tube Recirculation loop inlet* 10 mm OD tube Electrical Requirements Mains input...
All manuals and user guides at all-guides.com MEDICA Pro 30/60/120 US Operator Manual ELGA WARRANTY/CONDITIONS OF SALE ELGA LabWater is a trading name of VWS (UK) Ltd. General Limited Warranty VWS (UK) Ltd. warrants the products manufactured by it against defects in materials and workmanship when used in accordance with applicable instructions for a period of one year from the date of shipment for the products.
Page 35
All manuals and user guides at all-guides.com ELGA MEDICA Pro 30/60/120 US Operator Manual defective in materials or workmanship within the warranty period, provided the customer notifies VWS (UK) Ltd. promptly of any such defect. The cost of labor for the first ninety (90) days of the above warranty period is included in the warranty;...
Service office visit the country list on our website http://www.elgalabwater.com or contact ELGA at the number above. ELGA® is the global laboratory water brand name of Veolia Water. ELGA®, PURELAB®, MEDICA® and CENTRA® are trademarks of ELGA LabWater. Page 34...
Need help?
Do you have a question about the ELGA MEDICA Pro 30 and is the answer not in the manual?
Questions and answers