Sievers
Boron
*
Standard Addition Tool 2.0
Operation Manual
DLM 70500-01 EN REV A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sievers Boron Standard Addition Tool 2.0 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Veolia Sievers Boron Standard Addition Tool 2.0

  • Page 1 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN REV A...
  • Page 2 An electronic version of the StAT 2.0 Operation Manual is available upon request. Contact Sievers Technical Support. Sievers* is a trademark of Veolia and may be registered in one or more countries. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 3: Table Of Contents

    Sievers Boron Ultra Analyzer ..................43 Sievers UPW Boron Analyzer ................... 43 Chapter 2. System Description ................45 System Specifications ......................45 Recommended Specifications ..................45 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 4 Install the “INLET” and “OUTLET” Sample Lines .......... 67 Step 5. Install the “BYPASS” Drain Line ..............68 Step 6. Prime the Columns .................... 69 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 5 Power On and Off the Standard Addition Tool ..............84 Powering On the Standard Addition Tool ..............84 Powering Off the Standard Addition Tool ..............84 Navigating the Firmware screens ..................84 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 6 Chapter 5. Running Protocols ................102 Overview ..........................102 Prerequisites ........................103 Flushing the StAT 2.0 ....................... 103 To Flush the StAT 2.0 ..................104 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 7 To Prepare the StAT 2.0 for shipping ..............120 Resume Analysis on the Boron Analyzer ............... 125 Chapter 6. Maintenance ..................126 Overview ..........................126 Cleaning the StAT 2.0 ....................... 126 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 8 Step 4. Contact Sievers Technical Support ............... 132 Appendix A. Dimensions and Locations ............133 Overview ..........................133 Dimensions ........................134 Front View ....................... 134 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 9 Column Rinse Down Tool ............... 150 Overview ..........................150 Column Rinse Down Tool ....................150 Appendix D. Glossary ................... 152 Overview ..........................152 Glossary ..........................152 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 10: List Of Figures

    Install the Quick Disconnect Union Assembly ..............122 Figure 47. Draining the Mixing Tank ....................123 Figure 48. Connect Hand Pump Assembly to Column Fitting Tube Assembly ........ 124 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 11 Ethernet Cable Bracket on inner door below Touchscreen hardware ......149 Figure 67. StAT 2.0 Sample Connection Diagram using the Column Rinse Down Tool ....151 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 12: List Of Tables

    Flush Options Settings ...................... 93 Table 7. Output Concentrations by Protocol ................. 110 Table 8. Estimated Durations of a Full Syringe Injection Cycle ............ 111 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 13: Revision History

    DLM 70500-01 EN Rev A First Publication Mar 2025 RADEMARKS Sievers* is a trademark of Veolia and may be registered in one or more countries. John Guest is a registered trademark of John Guest International Limited; Swagelok is a ®...
  • Page 14: Warranty Service

    Telephone support is provided for troubleshooting and determination of parts to be shipped from Veolia to the customer in order to return the product to operation. If telephone support is not effective, the product may be returned to Veolia for repair or replacement. In some instances, suitable instruments may be available for short duration loan or lease.
  • Page 15 Operation and Maintenance Manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present. In no event shall Veolia be liable for incidental or consequential damages in connection with or arising from the use of the manual and its accompanying related materials.
  • Page 16: Warnings

    To avoid injury, keep hands away from the syringe while it is operating. If this Standard Addition Tool is used in a manner not specified by Veolia, the protection provided by the hardware may be impaired.
  • Page 17 Standard Addition Tool. Use of the supplied AC power cord will ensure safe operation of the Standard Addition Tool. Use of a power cord not supplied by Veolia is not authorized and may impair the safety of the equipment.
  • Page 18 Important! Do not attempt to access or edit any system parameters. Do not access or connect to the Firmware via any external devices without express permission from Sievers. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 19 情况下在标准品添加工具内部进行操作(包括安装维护部件),可能会导致人员受 伤。 瓶口含有锐利的针,设计用于刺穿样本瓶的隔片。请勿将手指或不当物体放入瓶口 中。 请勿将手放在注射器组件上。注射器由移动部件控制,会夹伤皮肤。 当注射器运行时,手应远离注射器,以免受伤。 如果未以Veolia公司规定的方法使用标准品添加工具,硬件提供的保护功能就可能会 被减弱。 作为常规运行的一部分,标准品添加工具在运行时使用标准品。在操作时,请始终使 用适当的个人防护设备(PPE)并遵循安全数据表(SDS)。 小心!在打开和关闭标准品添加工具的门时可能会夹到您的手!请将手拿开! 打开标准品添加工具的样品流之前,请先确保完成所有的管子连接,否则会使液体流 出,并导致硬件故障。 标准品添加工具里有混合罐,当标准品添加工具运行时,混合罐处于压力之下。标准 品添加工具的压力调节到 207 kPa (30 psig)。请检查压力容器是否损坏。如果损 坏,请勿运行标准品添加工具。 装有标准品添加工具的包装箱重达40磅(18 kg)以上。移动包装箱时需要两个人将 其抬起。Veolia建议由两个人抬起包装箱。 抬起和移动标准品添加工具时,只可使用它的把手,或抬起它的金属外壳。 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 20 这是安全等级I类产品,必须连接到接地电源。为了避免损坏标准品添加工具或造成 人员伤害,请确保将其连接到正确电压的电源线上。 本产品是A级产品。在室内环境中,本产品可能产生电磁干扰,此时用户需要采取适 当措施来消除干扰。 请参阅标准品随附的安全数据表(SDS),以确定标准品的正确使用、储存和处置 技术。 在关闭标准品添加工具前,必须先停止运行功能选项。 请非常小心,不要使电气元件受潮。 必须由合格的技术人员来执行所有的维护程序。 为了确保仪器安全工作,交流电源接地线必须牢固地连接到用螺栓固定在机壳上的安 装板上。 在安装电源线之前,请确保标准品添加工具侧面的电源开关处于关闭的位置。 任何需要打开标准品添加工具的电气组件盖子的操作(包括安装维护部件)都可能使 操作者受伤。为了避免潜在的危险电击,在打开标准品添加工具之前,请先关闭它的 电源开关,并尽可能断开它的电源线连接。 使用Veolia提供的交流电源线能够确保标准品添加工具的安全工作。不可使用非 Veolia 提供的电源线,否则可能会损害设备的安全性。 为了避免潜在的危险电击,在目视检查标准品添加工具时,请勿触摸里面的电气组件 盖子后面的任何部件。 为了避免人员受伤并确保设备准确运行,在标准品添加工具日常运行期间,请确保关 闭并锁紧它的门。 为了保护标准品添加工具免受损坏,在装运前请务必将标准品添加工具中的液体排 出。 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 21 警告 在清洁标准品添加工具时,必须将清洁液直接涂抹在抹布上,请勿将清洁液直接喷洒 在标准品添加工具上。在清洁标准品添加工具之前,请确保先关闭它的电源开关。请 勿用水或清洁液直接清洁触摸屏,因为水会损坏偏振器,或使其褪色。请立即用干净 的麂皮布或湿润的纤维素海绵擦去触摸屏上的冷凝水,以免触摸屏上出现水渍。 当样品流过标准品添加工具时,请勿移除试剂柱,因为此时液体会持续流过系统。 在标准品添加工具里面使用工具时,请务必小心。金属工具接触标准品添加工具内部 组件时可能会造成损坏。 在标准品添加工具里面的电气组件盖子后面进行操作时,请务必佩戴静电放电保护 (ESD)装备。 在将电源线连接到主电源之前,请先将电源线连接到标准品添加工具。 在安装电气接头时,请先断开标准品添加工具的所有电源连接。 交流电源线的墙上插头是标准品添加工具的主要电源断开装置。接通电源后,标准品 添加工具里的电气组件盖子后面会有高电压。为了避免电击风险,在拆下电气组件盖 子并在标准品添加工具里进行操作之前,请先断开标准品添加工具的电源连接。只有 在用标准品添加工具侧面的电源开关关闭电源并从交流电源插座上拔下主插头之后, 才可将电源线从标准品添加工具上拔下。 标准品添加工具上的此符号表示此产品符合中国 RoHS II 标准。 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 22: Français

    Pour éviter toute blessure, gardez les mains loin de la seringue lorsqu’elle est utilisée. Si cet outil d’ajout d’étalons est utilisé d’une manière non spécifiée par Veolia, la protection fournie par le matériel peut être altérée. L’outil d’ajout d’étalons utilise des étalons dans le cadre de son fonctionnement normal.
  • Page 23 Pour garantir un fonctionnement sûr de l’outil d’ajout d’étalons, utilisez le cordon d’alimentation fourni. L’utilisation d’un cordon d’alimentation non fourni par Veolia n’est pas autorisée et peut nuire à la sécurité de l’appareil. Afin d’éviter toute décharge potentiellement dangereuse, ne touchez à rien derrière le couvercle électronique à...
  • Page 24 CA. La présence de ce symbole sur l’outil d’ajout d’étalons indique que le produit est conforme à la directive RoHS II Chine. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 25: Deutsch

    Teile gesteuert, die die Haut einklemmen können. Um Verletzungen zu vermeiden, halten Sie die Hände während des Betriebs von der Spritze fern. Wird dieses Standard Addition Tool in einer von Veolia nicht spezifizierten Weise verwendet, kann der vom Gerät bereitgestellte Hardwareschutz beeinträchtigt werden.
  • Page 26 Stromversorgung, bevor Sie das Standard Addition Tool öffnen. Die Verwendung des mitgelieferten Netzkabels gewährleistet einen sicheren Betrieb des Standard Addition Tool. Die Verwendung eines nicht von Veolia gelieferten Netzkabels ist nicht zulässig und kann die Sicherheit des Gerätes beeinträchtigen.
  • Page 27 Seite abgezogen und der Netzstecker aus der Steckdose gezogen werden. Dieses Symbol auf dem Standard Addition Tool zeigt an, dass das Produkt die Anforderungen der China RoHS II erfüllt. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 28: Italiano

    Il container di spedizione per lo Standard Addition Tool pesa più di 40 libbre (18 kg ) ed è quindi necessaria una tecnica di sollevamento da parte di due persone durante la sua movimentazione. Veolia raccomanda di utilizzare due persone per sollevarlo.
  • Page 29 L'utilizzo del cavo di alimentazione elettrica a CA fornito in dotazione garantisce il funzionamento in sicurezza dello Standard Addition Tool. È vietato utilizzare un cavo di alimentazione elettrica non fornito da Veolia perché questo potrebbe pregiudicare la sicurezza dell'apparecchiatura. Per evitare scosse potenzialmente pericolose, non toccare niente dietro il coperchio dell'elettronica all'interno dello Standard Addition Tool quando si esegue un'ispezione visiva.
  • Page 30 CA. Questo simbolo sullo Standard Addition Tool indica che il prodotto è conforme alla direttiva RoHS II Cina. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 31: 日本語

    ます。 標準液添加ツールは、標準液添加ツールの運転中に圧力下で使用される混合タ ンクを含みます。標準液添加ツールは 207 kPa (30 psig) に調整されます。圧力 容器に損傷が無いか点検してください。損傷がある場合、標準液添加ツールを 運転しないでください。 標準液添加ツールの輸送用容器は 40 ポンド (18 kg) を超える重量があります - 輸送用容器を取り扱う際には、2 人で行う持ち上げ方法が必要です。Veolia で は、2 人で持ち上げることをお勧めします。 標準液添加ツールの持ち上げと運搬は、ハンドルまたは金属製筐体のみで装置 を取り扱って行う必要があります。 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 32 標準液添加ツールの電源を切る前に、必ず機能を停止させてください。 電気部品が濡れないように特に注意してください。 すべての保守作業を有資格の技術者が行う必要があります。 装置の安全な操作を確保するために、筐体にボルト留めされた取付プレートに しっかりと接続する必要があります。 電源コードの設置を行う前に、標準液添加ツールの側面にある電源トグル スイ ッチを必ず切るようにしてください。 保守品目の取り付けを含み、標準液添加ツール内の電子部品カバーの取り外し を必要とするいかなる操作も、人身傷害につながる可能性があります。潜在的 に危険な感電を避けるために、標準液添加ツールを開く前に電源を切り、可能 であれば、電源から切り離してください。 付属の AC 電源コードを使用することで、標準液添加ツールの安全な操作を確 保します。Veolia が提供していない電源コードの使用は認められず、機器の安 全性を損なう可能性があります。 潜在的に危険な感電を避けるため、目視検査中は標準液添加ツール内の電子部 品カバー裏のいかなる物にも触れないでください。 怪我を避けるため、そして正確な機能を確保するために、標準液添加ツールの ドアを必ず閉め、通常運転中には掛け金を掛けるようにしてください。 標準液添加ツールを損傷から守るために、輸送前に標準液添加ツールから液体 を必ず抜いてください。 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 33 す。 電子部品カバーの裏の標準液添加ツール内部で作業している間は、静電放電保 護を必ず使用してください。 主電源にコードを接続する前に、標準液添加ツールに電源コードを取り付けま す。 電気接続部の作業を行う場合には、標準液添加ツールへのすべての電源を切り 離してください。 AC 電源コードの壁プラグが、標準液添加ツール用の主な遮断装置です。電源が 接続されている場合、標準液添加ツールの電子部品カバー裏に高電圧が存在し ます。感電の危険性を排除するために、電子部品カバーが取り外された状態で 標準液添加ツール内部で作業する前に、標準液添加ツールへの電源を切ってく ださい。側面にある電源トグル スイッチを用いて標準液添加ツールの電源を切 り、壁の AC コンセントからメイン プラグを切り離して初めて、標準液添加ツ ールから電源コードを取り外す必要があります。 標準液添加ツールのこの記号は、製品が中国 RoHS II に準拠していることを示 します。 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 34: 한국어

    피부가 찝힐 수 있습니다. 부상을 예방하려면 주사기가 작동하는 동안 주사기에서 손을 떼십시오. 표준 추가 도구를 Veolia에서 규정하지 않은 방식으로 사용할 경우, 하드웨어가 제공하는 안전 보호 장치가 손상될 수 있습니다. 표준 추가 도구는 일반 작업의 일부로 표준을 사용합니다. 항상 적절한 개인 보호...
  • Page 35 피하려면 표준 추가 도구를 열기 전에 전원을 끄고 가능하면 전원 공급 장치에서 분리하십시오. 제공된 AC 전원 코드를 사용하면 표준 추가 도구를 안전하게 작동할 수 있습니다. Veolia에서 제공하지 않은 전원 코드의 사용은 승인되지 않았으며 장비의 안전이 손상될 수 있습니다. 위험한 충격이 발생하는 것을 방지하려면, 시각 검사 중에 표준 추가 도구 내부의...
  • Page 36 스위치로 표준 추가 도구의 전원을 끄고 AC 벽면 콘센트에서 메인 플러그를 분리한 후에만 표준 추가 도구에서 전원 코드를 분리해야 합니다. 표준 추가 도구에 있는 이 기호는 해당 제품이 중국 RoHS II를 준수함을 나타냅니다. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 37: Português

    Para evitar lesões, mantenha as mãos afastadas da seringa enquanto ela estiver em operação. Se a Ferramenta de Adição de Padrões for usada de maneira não especificada pela Veolia, a proteção fornecida pelo gabinete do aparelho poderá ser comprometida. A Ferramenta de Adição de Padrões usa padrões como parte da operação normal.
  • Page 38 Usar o cabo de alimentação CA fornecido garantirá a operação segura da Ferramenta de Adição de Padrões. Usar um cabo de alimentação não fornecido pela Veolia não é permitido e pode afetar a segurança do equipamento. Para evitar choque potencialmente perigoso, não toque em nada por trás da tampa dos componentes eletrônicos no interior da Ferramenta de Adição de...
  • Page 39 CA na parede. Este símbolo na Ferramenta de Adição de Padrões indica que o produto está em conformidade com a RoHS II da China. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 40: Español

    Si esta Herramienta de adición estándar se utiliza de una manera que no sea la recomendada por Veolia, puede que la protección provista por el hardware se vea reducida. La Herramienta de adición estándar emplea estándares como parte del funcionamiento normal.
  • Page 41 El uso del cable de alimentación CA suministrado garantiza el funcionamiento sin riesgo de la Herramienta de adición estándar. El uso de un cable de alimentación que no sea el suministrado por Veolia está prohibido y puede perjudicar la seguridad del equipo.
  • Page 42 El presente símbolo en la Herramienta de adición estándar indica que el producto cumple con la directriz relativa a las Restricciones a Sustancias Peligrosas versión II de China. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 43: Chapter 1. Introduction

    Chapter 1. Introduction Theory of Operation The Sievers Boron Standard Addition Tool 2.0, also referred to as “Standard Addition Tool” and “StAT 2.0” within this document, is designed to take a known Standard concentration and dilute it into a controlled concentration of Standard Solution and deliver that solution to a connected Boron Analyzer.
  • Page 44 Boron Analyzer prior to initiating a protocol. Refer to the individual Analyzer’s Operation and Maintenance Manual for more information. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 45: Chapter 2. System Description

    Sievers recommends using a slower Sample Flow Rate when performing protocols. For more information see “Sievers Recommended Sample Flow Rate Ranges, by protocol” on page 81. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 46: System Overview

    It is not recommended to perform Protocols with less than 25 psig due to the internal drop in pressure within the StAT 2.0. The UPW Boron Analyzer requires 20 psig sample pressure at the Analyzer SAMPLE INLET to be able to perform analysis. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 47: Boron Ultra Analyzer

    • Major Subsystems The Standard Addition Tool has two major subsystems. See the following: Fluidics Subsystem on page • Electronics Subsystem on page • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 48: Fluidics Subsystem

    Path mix before entering the Mixing Tank. The top of the Mixing Tube connects to the Syringe Assembly. The bottom of the Mixing Tube connects to the INLET of the Mixing Tank. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 49: Electronics Subsystem

    Terminal Blocks • PLC Controller • Touchscreen • Power Entry Module; including Power Toggle Switch NOTE: There is an internal fuse within this component. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 50: Flow Paths

    Standard Flow Path, perform a Dead Volume Flush. For more information, refer to “Dead Volume Flush Flow Path” on page 54. Syringe Flow Paths and Diagrams This section describes the different Syringe Flush settings. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 51: Fill From Vial

    Vial Port. See Figure The fluid will flow from the Stream to the V2 Valve, then the V1 Valve, and then • finally into the Syringe. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 52: Empty To Waste

    This is used primarily when priming the Syringe with Standard. Emptying to the WASTE container ensures the Syringe flushes with Standard without injecting the Standard into the stream and causing a large spike in boron values. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 53: Empty To Stream

    Standard within the tubing. Primarily, this option is used when preparing the unit for shipping to purge the dead volume with air leading to the Mixing Point. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 54: Dead Volume Flush Flow Path

    This flow path within the StAT 2.0 diverts some of the sample flow OUT through the “BYPASS” Port to a gravity drain to achieve a higher Sample Flow Rate. Use this to Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 55 “Sample Flow Rate (mL/min)” range for the protocol being performed. For more information on using this option, see “Using the Bypass Flow Path” on page 77. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 56: Chapter 3. Installation

    Refer to the Boron Analyzer Operation and Maintenance Manual, as needed. If the Boron Analyzer is performing measurements, stop analysis. • Boron Ultra Analyzer Only. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 57: Step 1. Unpack The Standard Addition Tool

    “Table 1:Accessories Kit Packing List” on page 61, as needed. Boron Ultra Analyzer Only. Refer to the Operation and Maintenance Manual for more information. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 58: Figure 5. Accessories Kit And Foam Pieces

    If needed, reconnect it to the Ethernet Port. Ensure the cable is secured through the cable bracket. Refer to Figure 6 on page 59, as needed. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 59: Figure 6. Ethernet Cable Bracket On Inner Door Below Touchscreen

    Open the Shipping Container while it is still vertical and carefully tip the StAT 2.0 enclosure out of the shipping foam pieces using the front handle on the top of the StAT Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 60: Figure 8. Remove The Enclosure, Option 2

    Column has two plugs sealing the Column Block Ports. Leave the plugs in place until the Column is installed to prevent contamination. The power cord will be stored inside the shipping container tucked in alongside the enclosure. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 61: Table 1. Accessories Kit Packing List

    Extra Parts: Quick Disconnect Waste Fitting (Qty • Use this tool when draining the • Mixing Tank. FITTING, 1/4", JG, ELBOW (Qty 6) • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 62 This union will be used when • draining the StAT 2.0. Waste Container (Qty 1) — • This container will collect any Syringe waste created during protocols and flushes. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 63: Step 2. Select A Location For The Standard Addition Tool

    StAT 2.0 Dimensions Width: 52.3 cm (20.6 in) Height: 65.8 cm (25.69 in) Depth: 33.3 cm (13.1 in) Door Swing 50.8 cm (20 in) front clearance: Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 64: Recommended Service Area

    Fully opening the enclosure front door to allow for full access to the unit’s interior. • Allowing room for a Waste Container to be installed nearby. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 65: Step 3. Install The Stat 2.0 "Waste" Line

    Important! Ensure the same section of tubing is used as a Waste Line for each subsequent installation of this unit until it needs to be replaced. Do not connect a Waste tube section to any other Port on the StAT 2.0. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 66: Figure 10. External Port Connections

    Installation Install the StAT 2.0 “WASTE” Line External Port Connections Figure 10. Install the “WASTE” Line Figure 11. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 67: Step 4. Install The "Inlet" And "Outlet" Sample Lines

    68, as needed. For the Boron Ultra Analyzer, this will always be “SAMPLE INLET 1.” • Important! Sample flow should remain OFF at this time. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 68: Step 5. Install The "Bypass" Drain Line

    Port arrives plugged. Remove the plug and set it aside somewhere safe. Do not lose this hardware! The “BYPASS” Port connection uses a John Guest (JG) fitting. Refer to Figure 10 on page 66, ® as needed. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 69: Step 6. Prime The Columns

    StAT 2.0. If needed, refer to Figure 57 Overall Interior Locations” on page 141 for a more detailed view. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 70: Figure 14. Internal Locations

    • Manifold behind the Locking Pins). See Figure Column Holder (bracket located below the Column Manifold that holds the Column • tubes in place.) Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 71: Figure 15. Tube Stems

    Block Port above the connected tube. Carefully fill the Column until it is completely full to displace as much air as possible. See Figure Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 72: Figure 16. Fill The Columns With Upw

    Tube Stems. Refer to Figure 17 Figure 18 as needed. This can be difficult. Using the Locking Pin Figure 17. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 73: Figure 18. Install The Columns

    Shipping Bracket to the rear of the enclosure. Remove the two Thumb Screws on the front of the Shipping Brackets in front of the Mixing Tank. Remove the Shipping Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 74: Figure 19. Using The Riser Block

    Mixing Tank. To prevent damage, do not attempt to close the door! Outlet Fitting and Tubing Mixing Tank Riser Block Using the Riser Block Figure 19. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 75: Step 8. Connect To Power

    Power Cycle the unit if there are any startup alarms or if the unit will not “Home” the Syringe. If needed, power cycle the unit multiple times. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 76 Important! The StAT 2.0 operating pressure limit is < 35 psig. Do not increase the Pressure Regulator beyond 35 psig to prevent internal leaking. Customer sample supply inlet pressure to the StAT 2.0. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 77 StAT 2.0 out to a vented gravity drain through the “BYPASS” Port. This will increase the Sample Flow Rate. The Sample Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 78 Remove the remaining air by carefully moving/agitating/shaking the Mixing Tank back and forth on an incline to allow any remaining air bubbles to leave via the Mixing Tank Outlet Tubing. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 79: Figure 21. Compressing The Mixing Tank Outlet Tubing

    Compressing the Mixing Tank Outlet tubing Figure 21. The Mixing Tank is considered primed when the following statements are true: Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 80 Start Online analysis using the shortest Cycle Duration available (3-minute). Continue performing Online analysis until the measured Baseline Average boron value from the connected Boron Ultra Analyzer consistently displays ≤ 0.020 ppb Boron. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 81 Pressure Regulator beyond 35 psig to prevent internal leaking. It is possible to perform Protocol(s) outside of the recommended Sample Flow Rate Range and Sample Pressure Ranges. This manual describes using the recommended specifications. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 82 Standard Solution delivery, and reduce the time available to run a protocol. For more information, see “Table 7: Estimated Durations of a Full Syringe Injection Cycle” on page 110. Boron Ultra Analyzer Only. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 83: Step 16. Installation Complete

    Firmware screens. When ready, refer to “Chapter 5: Running Protocols” on page 102 for detailed information about preparing the StAT 2.0 to run protocol(s). Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 84: Chapter 4. Basic Operation

    The Firmware contains three primary Menu screens: Main Menu, Manual Menu, and Setup Menu. Use the three high level tab buttons on the top of the Touchscreen to navigate between the main menus. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 85: Table 3. Text Field Types

    User to navigate away from the current screen. To view a different screen either wait for the active process to complete (if applicable) or “Cancel” the active process. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 86: Figure 22. Using The Numeric Keypad

    Important! All “Syringe Positions (steps)” values shown in this manual are representative and not intended to provide specific direction during any procedures or protocols. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 87: Figure 23. Home Syringe Button Overlay

    To manually initiate a “Syringe Homing” sequence, see “HOME SYRINGE” on page 100. These values are provided as examples only. The actual “Syringe Position (steps)” values displayed may differ. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 88: Figure 24. Main Menu → Auto Deliver

    The AUTO DELIVERY function prepares the Standard Solution for an addition or injection in preparation for running a protocol on the connected Boron Analyzer. For more information, see the “StAT 2.0 Startup Procedure” on page 106. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 89: Table 4. Auto Delivery Settings

    Analyzer. This value is protocol dependent. • This value is also used to calculate the “Boron Reference Concentration” when performing a protocol. • Minimum: 0 Maximum: 999999999 • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 90: Figure 25. Main Menu → Fixed Deliver

    The FIXED DELIVERY function is a marginal utility for troubleshooting. • The tool allows the User to START and STOP the Fixed Delivery Function. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 91: Table 5. Fixed Delivery Settings

    Function Fixed Deliver Rate (µL/min): Input the value of the desired Fixed Deliver Rate. • • Default: 0.3 Minimum: 0.1 • • Maximum: 500 Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 92: Figure 26. Main Menu → Flush

    Press the FLUSH START button to perform the Syringe Flush sequence using the visible settings. Once pushed, the button will change color and display new text “FLUSH STARTED” to indicate that it is active. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 93: Figure 27. Flush Started (Example Settings Shown Selected)

    Port and into the Waste Container. The Syringe Plunger will move UP during this step. • Refer to Figure 3 on page 53 for a Flow Diagram. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 94: Figure 28. Main Menu → Dead Vol Flush

    These settings are based on the volume of the Syringe Assembly and the length of the tubing connections. The Dead Vol Flush Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 95: Manual Menu

    Manual Menu The Manual Menu has two tabs. Manual Menu → Semi Manual on page 96. • Manual Menu → Manual on page 97. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 96: Figure 30. Manual Menu → Semi Manual

    This function commands the Syringe Assembly to • push the Syringe contents out to the “WASTE” Port. The Syringe Plunger will move UP. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 97: Figure 31. Manual Menu → Manual

    These functions will only be active while the button is pressed. A Short Press will activate the function once. A Long Press will continuously perform the selected process while the button is pressed. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 98: Setup Menu

    Setup Menu → Flow Meter on page 99. • Setup Menu → Syringe on page 100. • Setup Menu → Engineering on page 101. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 99: Figure 32. Setup Menu → Flow Meter

    The values displayed on the Flow Meter tab are specific to the hardware installed. Any component replacement requires verification of the other StAT 2.0 hardware and this task is only performed at the Factory. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 100: Figure 33. Setup Menu → Syringe

    “Syringe Position (steps)” value stops changing in the bottom left corner of the Touchscreen. NOTE: Not all screens display the “Syringe Position (steps)” value in the bottom left corner of the Touchscreen. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 101: Exporting Data

    Use the connected Boron Analyzer to export any required Verification History or Calibration History, if applicable. Consult the Operation and Maintenance Manual for the connected Boron Analyzer model. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 102: Chapter 5. Running Protocols

    Sievers Boron Ultra Analyzer and a Sievers UPW Boron Analyzer. Follow the instructions for the appropriate Boron Analyzer model. Boron Ultra Analyzer Only. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 103 PPE and follow SDS guidelines. The vial port contains sharp needles designed to pierce the septa of sample vials. Do not put fingers or inappropriate materials into the vial port. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 104: Figure 35. Flush Options, Stream Setup

    9. From the Flush Options: Ensure the follow settings are visible: FILL FROM STREAM • EMPTY TO WASTE • SYRINGE EMPTY • Flush Options, STREAM Setup Figure 35. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 105: Figure 36. Flush Options, Vial Setup

    14. Reduce or increase the Sample Flow Rate to be within the recommended optimal recommended protocol range, if needed. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 106 Empty Waste Container; installed at the end of the StAT 2.0 gravity “WASTE” line. • StAT 2.0 Startup Procedure Important! The StAT 2.0 must be flushed and a new Baseline Average established between protocols. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 107 Standard Vial and flush it out to the “WASTE” line into the Waste Container. This Flush process primes the Syringe Assembly tubing with the Standard to replace any previous fluid that may have been in the lines. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 108: Figure 37. Flush Options, Fill Syringe With Standard

    B. Press the DEAD VOL FLUSH START button. C. Wait for the process to complete. The Dead Vol Flush Cycle takes 180 seconds to complete. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 109: Figure 38. Manual Menu → Semi Manual

    Important! Using a slower Sample Flow Rate will increase the expected duration of the Full Syringe Injection Cycle. For more information, see “Table 8: Estimated Durations of a Full Syringe Injection Cycle” on page 111. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 110: Figure 39. Main Menu → Auto Deliver (Example Settings Shown)

    0.05 ppb (50 ppt) UPW Boron Analyzer Verification 2.0 ppb NOTE: The other settings on this screen will be populated by the Firmware. • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 111: Table 8. Estimated Durations Of A Full Syringe Injection Cycle

    These Sample Flow Rates are provided as examples only. These values demonstrate the difference in how changing the Sample Flow Rate will affect the estimated duration of a Full Syringe Injection Cycle. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 112: Calibration Protocol

    Standard Solution concentration threshold met. For more information see “StAT 2.0 Startup Procedure” on page 106. Boron Ultra Analyzer Only. Boron Ultra Analyzer Only. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 113: Figure 40. Maintenance → Calibration / Verification (Calibration Selected)

    Main Menu → Auto Deliver → Output Concentration (PPB) Calculate the Baseline Average from the last hour of boron measurements from the connected Boron Analyzer. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 114 The Boron RSD is 3% or less. Under most circumstances, do not apply a • calibration if it fails due to high RSDs. High RSDs could indicate inconsistent flow or Standard Delivery issue. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 115 Ultra Analyzer and a Sievers UPW Boron Analyzer. Follow the instructions for the correct Boron Analyzer. Verifying the Boron Ultra Analyzer Use this section to Verify a Sievers Boron Ultra Analyzer. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 116: Figure 41. Maintenance → Calibration / Verification (Verification Selected)

    Main Menu → Auto Deliver → Output Concentration (PPB) Calculate the Baseline Average from the last hour of boron measurements from the connected Boron Analyzer. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 117 Follow your internal service procedures and follow the prompts on the Sievers UPW Boron Analyzer to complete the Verification procedure. Refer to your internal service procedures to understand the Verification procedure results to understand the Pass/Fail Criteria. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 118 A. From the StAT 2.0 Touchscreen, navigate to the following screen: Main Menu → Flush B. Select the following settings: Number of Cycles: 20 • Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 119: Figure 42. Flush Options For Shut Down; Stream

    Flush Options: (toggle the buttons until the following settings are visible) • FILL FROM VIAL • EMPTY TO WASTE • SYRINGE EMPTY • Flush Options for Shut Down; VIAL Figure 43. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 120: Equipment Needed

    “Resume Analysis on the Boron Analyzer” on page 125. 6. From the StAT 2.0 Touchscreen, navigate to the following screen: Manual Menu → Semi Manual Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 121: Figure 44. Manual Menu → Semi Manual

    C. Disconnect the Mixing Tank INLET and OUTLET Quick Connect Fittings. Connect the two tube ends together using the using the provided Quick Disconnect Union Assembly. See Figure Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 122: Figure 46. Install The Quick Disconnect Union Assembly

    Tank from the enclosure. Using one hand, hold on to the Mixing Tank Handle at the rear of the Mixing Tank and use the other hand to support the front of the Mixing Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 123: Figure 47. Draining The Mixing Tank

    Use a long 4 mm T-Handle Hex Wrench to attach the two hex screws to the rear wall of the enclosure. Secure the front of the Shipping Brackets in place using the two Thumb Screws that were removed during installation. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 124: Figure 48. Connect Hand Pump Assembly To Column Fitting Tube Assembly

    D. Rotate the Hand Pump Vent Valve closed (clockwise), if not already done. E. Use the Hand Pump Assembly to push air through each Column until it is dry. Perform at least 10 compressions of the squeeze bulb. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 125: Figure 49. Reinstall The External Port Plugs

    Restore the Boron Analyzer to the previous analysis settings that were recorded earlier. Ensure to re-enable any disabled Streams, if applicable. Perform a Backup, if available, or record the new settings and then resume analysis. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 126: Chapter 6. Maintenance

    Consumables can be purchased from the Sievers website or contact Sievers Technical Support for more information. Standards Procure any required Calibration or Verification Standards from Sievers Technical Support. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 127: Calibration And Verification

    Sievers. Contact Sievers Technical Support for more information. For more information, see page 2. To locate the current firmware/software version, from the StAT 2.0, navigate to the following screen: Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 128 Important! Do not attempt to access or edit any system parameters. Do not access or connect to the Firmware via any external devices without express permission from Sievers. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 129: Chapter 7. Troubleshooting

    Step 2. Review Alarms and Errors If any alerts (Alarms or Errors) occurred while operating the StAT 2.0, consult the “Appendix B:Alarms and Errors” on page 145. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 130 The customer supply line Shutoff Valve is OPEN. • The StAT 2.0 “INLET” Port connection is properly connected on the left side of the • enclosure. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 131 The Ethernet Cable connection on the inside of the door can loosen after the unit has been shipped. Check, and if needed, fully reconnect the Ethernet Cable connection into the Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 132 Step 4. Contact Sievers Technical Support If Basic Troubleshooting steps do not remedy the issue, please contact Sievers Technical Support. For more information, see page 2. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 133: Appendix A. Dimensions And Locations

    Dimensions and Locations Overview This section contains illustrations of the Analyzer dimensions and locations with callouts. • Dimensions on page 134. • Locations on page 138. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 134: Figure 50. Front View Dimensions

    Dimensions and Locations Dimensions Dimensions Front View Front View Dimensions Figure 50. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 135: Figure 51. Top View Dimensions

    Dimensions and Locations Dimensions Top View Top View Dimensions Figure 51. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 136: Figure 52. Left Side View Dimensions

    Dimensions and Locations Dimensions Left Side View Left Side View Dimensions Figure 52. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 137: Figure 53. Right Side View

    Dimensions and Locations Dimensions Right Side View Right Side View Figure 53. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 138: Figure 54. Front View Locations

    Dimensions and Locations Locations Locations External Locations Front Locations Front View Locations Figure 54. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 139: Figure 55. Left Side Locations

    Dimensions and Locations Locations Left Side Locations Left Side Locations Figure 55. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 140: Figure 56. Right Side Locations

    Dimensions and Locations Locations Right Side Locations Right Side Locations Figure 56. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 141: Figure 57. Overall Interior Locations

    Dimensions and Locations Locations Interior Locations Overall Interior View Figure Figure Overall Interior Locations Figure 57. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 142: Figure 58. Column Assembly Locations

    Dimensions and Locations Locations Column Assembly Column Assembly Locations Figure 58. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 143: Figure 59. Left Side Internal Locations

    Dimensions and Locations Locations Left side internal locations Left side internal locations Figure 59. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 144: Figure 60. Inner Door Locations

    Dimensions and Locations Locations Inner Door Locations Inner Door Locations Figure 60. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 145: Appendix B. Alarms And Errors

    Sample Flow Rate, but large fluctuations can cause errors. This change in Sample Flow Rate can occur if there is loss in line pressure from a loose or Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A ©...
  • Page 146: Figure 61. Flow Sensor Deviation Alarm

    Alarm. If needed, power cycle the unit multiple times. If this does not resolve the issue, consult “Chapter 7 Troubleshooting” on page 129. Homing Timeout Alarm Figure 62. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 147: Figure 63. Example "Numeric Parameter Out Of Specification Error

    1. The firmware must first be displaying a “Numeric Parameter Out of Specification Error.” For an example, see Figure 63 on page 147. Example “Numeric Parameter Out of Specification Error” Figure 63. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 148: Figure 64. Example Error Window

    It may take a few minutes for the firmware to reconnect and refresh the interface. If needed, power cycle the unit. For more information, see “Power On and Off the Standard Addition Tool” on page 84. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 149: Figure 65. Example Error State: "Plc Communication Error

    Alarms and Errors Errors Example Error State: “PLC Communication Error” Figure 65. Ethernet Cable Bracket on inner door below Touchscreen hardware Figure 66. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 150: Appendix C. Column Rinse Down Tool

    Reagent Column of the water before installing it into the Boron Ultra Analyzer! Failure to drain the water may result in diluted reagent in the Reagent Reservoir. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
  • Page 151: Figure 67. Stat 2.0 Sample Connection Diagram Using The Column Rinse Down Tool

    Column Rinse Down Tool Column Rinse Down Tool StAT 2.0 Sample Connection Diagram using the Column Rinse Down Tool Figure 67. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 152 Standard into the sample stream at a fixed rate, instead of a variable rate. This function is not used during when performing a protocol. Boron Ultra Analyzer Only. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...
  • Page 153 Standard Solution. Stream This term refers to the sample flow path within the StAT 2.0. Syringe Flush See “Flush.” Ultrapure Water. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A © 2025 Veolia...