DLM 70500-01 EN Rev A First Publication Mar 2025 RADEMARKS Sievers* is a trademark of Veolia and may be registered in one or more countries. John Guest is a registered trademark of John Guest International Limited; Swagelok is a ®...
Telephone support is provided for troubleshooting and determination of parts to be shipped from Veolia to the customer in order to return the product to operation. If telephone support is not effective, the product may be returned to Veolia for repair or replacement. In some instances, suitable instruments may be available for short duration loan or lease.
Page 15
Operation and Maintenance Manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present. In no event shall Veolia be liable for incidental or consequential damages in connection with or arising from the use of the manual and its accompanying related materials.
To avoid injury, keep hands away from the syringe while it is operating. If this Standard Addition Tool is used in a manner not specified by Veolia, the protection provided by the hardware may be impaired.
Page 17
Standard Addition Tool. Use of the supplied AC power cord will ensure safe operation of the Standard Addition Tool. Use of a power cord not supplied by Veolia is not authorized and may impair the safety of the equipment.
Page 18
Important! Do not attempt to access or edit any system parameters. Do not access or connect to the Firmware via any external devices without express permission from Sievers. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Pour éviter toute blessure, gardez les mains loin de la seringue lorsqu’elle est utilisée. Si cet outil d’ajout d’étalons est utilisé d’une manière non spécifiée par Veolia, la protection fournie par le matériel peut être altérée. L’outil d’ajout d’étalons utilise des étalons dans le cadre de son fonctionnement normal.
Page 23
Pour garantir un fonctionnement sûr de l’outil d’ajout d’étalons, utilisez le cordon d’alimentation fourni. L’utilisation d’un cordon d’alimentation non fourni par Veolia n’est pas autorisée et peut nuire à la sécurité de l’appareil. Afin d’éviter toute décharge potentiellement dangereuse, ne touchez à rien derrière le couvercle électronique à...
Teile gesteuert, die die Haut einklemmen können. Um Verletzungen zu vermeiden, halten Sie die Hände während des Betriebs von der Spritze fern. Wird dieses Standard Addition Tool in einer von Veolia nicht spezifizierten Weise verwendet, kann der vom Gerät bereitgestellte Hardwareschutz beeinträchtigt werden.
Page 26
Stromversorgung, bevor Sie das Standard Addition Tool öffnen. Die Verwendung des mitgelieferten Netzkabels gewährleistet einen sicheren Betrieb des Standard Addition Tool. Die Verwendung eines nicht von Veolia gelieferten Netzkabels ist nicht zulässig und kann die Sicherheit des Gerätes beeinträchtigen.
Page 27
Seite abgezogen und der Netzstecker aus der Steckdose gezogen werden. Dieses Symbol auf dem Standard Addition Tool zeigt an, dass das Produkt die Anforderungen der China RoHS II erfüllt. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Il container di spedizione per lo Standard Addition Tool pesa più di 40 libbre (18 kg ) ed è quindi necessaria una tecnica di sollevamento da parte di due persone durante la sua movimentazione. Veolia raccomanda di utilizzare due persone per sollevarlo.
Page 29
L'utilizzo del cavo di alimentazione elettrica a CA fornito in dotazione garantisce il funzionamento in sicurezza dello Standard Addition Tool. È vietato utilizzare un cavo di alimentazione elettrica non fornito da Veolia perché questo potrebbe pregiudicare la sicurezza dell'apparecchiatura. Per evitare scosse potenzialmente pericolose, non toccare niente dietro il coperchio dell'elettronica all'interno dello Standard Addition Tool quando si esegue un'ispezione visiva.
피부가 찝힐 수 있습니다. 부상을 예방하려면 주사기가 작동하는 동안 주사기에서 손을 떼십시오. 표준 추가 도구를 Veolia에서 규정하지 않은 방식으로 사용할 경우, 하드웨어가 제공하는 안전 보호 장치가 손상될 수 있습니다. 표준 추가 도구는 일반 작업의 일부로 표준을 사용합니다. 항상 적절한 개인 보호...
Page 35
피하려면 표준 추가 도구를 열기 전에 전원을 끄고 가능하면 전원 공급 장치에서 분리하십시오. 제공된 AC 전원 코드를 사용하면 표준 추가 도구를 안전하게 작동할 수 있습니다. Veolia에서 제공하지 않은 전원 코드의 사용은 승인되지 않았으며 장비의 안전이 손상될 수 있습니다. 위험한 충격이 발생하는 것을 방지하려면, 시각 검사 중에 표준 추가 도구 내부의...
Para evitar lesões, mantenha as mãos afastadas da seringa enquanto ela estiver em operação. Se a Ferramenta de Adição de Padrões for usada de maneira não especificada pela Veolia, a proteção fornecida pelo gabinete do aparelho poderá ser comprometida. A Ferramenta de Adição de Padrões usa padrões como parte da operação normal.
Page 38
Usar o cabo de alimentação CA fornecido garantirá a operação segura da Ferramenta de Adição de Padrões. Usar um cabo de alimentação não fornecido pela Veolia não é permitido e pode afetar a segurança do equipamento. Para evitar choque potencialmente perigoso, não toque em nada por trás da tampa dos componentes eletrônicos no interior da Ferramenta de Adição de...
Page 39
CA na parede. Este símbolo na Ferramenta de Adição de Padrões indica que o produto está em conformidade com a RoHS II da China. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Si esta Herramienta de adición estándar se utiliza de una manera que no sea la recomendada por Veolia, puede que la protección provista por el hardware se vea reducida. La Herramienta de adición estándar emplea estándares como parte del funcionamiento normal.
Page 41
El uso del cable de alimentación CA suministrado garantiza el funcionamiento sin riesgo de la Herramienta de adición estándar. El uso de un cable de alimentación que no sea el suministrado por Veolia está prohibido y puede perjudicar la seguridad del equipo.
Page 42
El presente símbolo en la Herramienta de adición estándar indica que el producto cumple con la directriz relativa a las Restricciones a Sustancias Peligrosas versión II de China. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Chapter 1. Introduction Theory of Operation The Sievers Boron Standard Addition Tool 2.0, also referred to as “Standard Addition Tool” and “StAT 2.0” within this document, is designed to take a known Standard concentration and dilute it into a controlled concentration of Standard Solution and deliver that solution to a connected Boron Analyzer.
It is not recommended to perform Protocols with less than 25 psig due to the internal drop in pressure within the StAT 2.0. The UPW Boron Analyzer requires 20 psig sample pressure at the Analyzer SAMPLE INLET to be able to perform analysis. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Path mix before entering the Mixing Tank. The top of the Mixing Tube connects to the Syringe Assembly. The bottom of the Mixing Tube connects to the INLET of the Mixing Tank. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
This is used primarily when priming the Syringe with Standard. Emptying to the WASTE container ensures the Syringe flushes with Standard without injecting the Standard into the stream and causing a large spike in boron values. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Standard within the tubing. Primarily, this option is used when preparing the unit for shipping to purge the dead volume with air leading to the Mixing Point. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Column has two plugs sealing the Column Block Ports. Leave the plugs in place until the Column is installed to prevent contamination. The power cord will be stored inside the shipping container tucked in alongside the enclosure. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Important! Ensure the same section of tubing is used as a Waste Line for each subsequent installation of this unit until it needs to be replaced. Do not connect a Waste tube section to any other Port on the StAT 2.0. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Power Cycle the unit if there are any startup alarms or if the unit will not “Home” the Syringe. If needed, power cycle the unit multiple times. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 76
Important! The StAT 2.0 operating pressure limit is < 35 psig. Do not increase the Pressure Regulator beyond 35 psig to prevent internal leaking. Customer sample supply inlet pressure to the StAT 2.0. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 77
StAT 2.0 out to a vented gravity drain through the “BYPASS” Port. This will increase the Sample Flow Rate. The Sample Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 78
Remove the remaining air by carefully moving/agitating/shaking the Mixing Tank back and forth on an incline to allow any remaining air bubbles to leave via the Mixing Tank Outlet Tubing. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 80
Start Online analysis using the shortest Cycle Duration available (3-minute). Continue performing Online analysis until the measured Baseline Average boron value from the connected Boron Ultra Analyzer consistently displays ≤ 0.020 ppb Boron. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 81
Pressure Regulator beyond 35 psig to prevent internal leaking. It is possible to perform Protocol(s) outside of the recommended Sample Flow Rate Range and Sample Pressure Ranges. This manual describes using the recommended specifications. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
The Firmware contains three primary Menu screens: Main Menu, Manual Menu, and Setup Menu. Use the three high level tab buttons on the top of the Touchscreen to navigate between the main menus. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
User to navigate away from the current screen. To view a different screen either wait for the active process to complete (if applicable) or “Cancel” the active process. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Press the FLUSH START button to perform the Syringe Flush sequence using the visible settings. Once pushed, the button will change color and display new text “FLUSH STARTED” to indicate that it is active. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
These settings are based on the volume of the Syringe Assembly and the length of the tubing connections. The Dead Vol Flush Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
These functions will only be active while the button is pressed. A Short Press will activate the function once. A Long Press will continuously perform the selected process while the button is pressed. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
The values displayed on the Flow Meter tab are specific to the hardware installed. Any component replacement requires verification of the other StAT 2.0 hardware and this task is only performed at the Factory. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Use the connected Boron Analyzer to export any required Verification History or Calibration History, if applicable. Consult the Operation and Maintenance Manual for the connected Boron Analyzer model. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 103
PPE and follow SDS guidelines. The vial port contains sharp needles designed to pierce the septa of sample vials. Do not put fingers or inappropriate materials into the vial port. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 107
Standard Vial and flush it out to the “WASTE” line into the Waste Container. This Flush process primes the Syringe Assembly tubing with the Standard to replace any previous fluid that may have been in the lines. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Use a long 4 mm T-Handle Hex Wrench to attach the two hex screws to the rear wall of the enclosure. Secure the front of the Shipping Brackets in place using the two Thumb Screws that were removed during installation. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
D. Rotate the Hand Pump Vent Valve closed (clockwise), if not already done. E. Use the Hand Pump Assembly to push air through each Column until it is dry. Perform at least 10 compressions of the squeeze bulb. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Restore the Boron Analyzer to the previous analysis settings that were recorded earlier. Ensure to re-enable any disabled Streams, if applicable. Perform a Backup, if available, or record the new settings and then resume analysis. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 128
Important! Do not attempt to access or edit any system parameters. Do not access or connect to the Firmware via any external devices without express permission from Sievers. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Page 131
The Ethernet Cable connection on the inside of the door can loosen after the unit has been shipped. Check, and if needed, fully reconnect the Ethernet Cable connection into the Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...
Reagent Column of the water before installing it into the Boron Ultra Analyzer! Failure to drain the water may result in diluted reagent in the Reagent Reservoir. Sievers Boron Standard Addition Tool 2.0 Operation Manual DLM 70500-01 EN Rev A...