Verathon GLIDESCOPE GO Operation & Maintenance Manual page 26

Table of Contents

Advertisement

ELECTROMAGNETIC COMPATIBILITY
The system is designed to be in compliance with IEC 60601-1-2, which contains
electromagnetic compatibility (EMC) requirements for medical electrical equipment.
The limits for emissions and immunity specified in this standard are designed to provide
reasonable protection against harmful interference in a typical medical installation.
The system complies with the applicable essential performance requirements specified
in IEC 60601-1 and IEC 60601-2-18. Results of immunity testing show that the essential
performance of the system is not affected under the test conditions described in the
following tables.
ELECTROMAGNETIC IMMUNITY
Table 4.
Guidance and Manufacturer's Declaration —Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified below. The customer or the
user of the system should ensure that it is used in such an environment.
Immunity Tests
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
(Repetition
frequency 100
kHz)
Surge transient
IEC 61000-4-5
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
20
0900-4788 REV-10
IEC 60601 Test Level
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Voltage dips:
• 0% during 0.5 cycle
• 0% during 1 cycle
• 40% during 10/12
cycles at 50/60 Hz
• 70% during 25/30
cycles at 50/60 Hz
Voltage interruptions:
• 0% during 250/300
cycles at 50/60 Hz
30 A/m; 50/60 Hz
Compliance
Electromagnetic Environment –
Level
Floors should be wood, concrete,
or ceramic tile. If floors are covered
In compliance
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
In compliance
of a home healthcare or hospital
environment.
Mains power quality should be that
In compliance
of a home healthcare or hospital
environment.
Mains power quality should be that
of a home healthcare or hospital
environment. If the user of the
system requires continued operation
In compliance
during power mains interruptions,
it is recommended that the system
be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of a
In compliance
typical location in a home healthcare
or hospital environment.
Guidance