Troubleshooting; Warranty - J.P. SELECTA RS 3001-V Manual

Evaporador rotativo
Table of Contents

Advertisement

Available languages

Available languages

MANUAL DE INSTRUCCIONES CODIGO 80370 REV 0
J
.P. SELECTA s.a.u.

10 TROUBLESHOOTING

Description
It does not heat
The cooling water is
not acting
It does not generate
vacuum
The sample flask ro-
tor does not rotate

11 WARRANTY

This product is guaranteed for one year. It does not cover damages caused by an
incorrect use or causes beyond the control of JP. Selecta, s.a.u.
Any manipulation of the equipment by personnel not authorized by JP. Selecta, s.a.u.
automatically cancels the guarantee.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
Juliol/2018
(Sujetas a modificaciones sin previo aviso)
Cause
Working temperature
lower than room tem-
perature
Resistance is melted
It has any element
which closes the pass
or rotation of water
The engine does not
work
The pump is not
switched on
The start-up switch is
stopped
There are leaks in the
hose joints or in the
flasks connection
The connection cable is
not connected
The startup switch is off Follow chapter "7.Opera-
The speed setting po-
tentiometer is at zero
Resolution
Check chapter "7.Opera-
tion" to select the temper-
ature
Contact with the technical
service for replacement
Remove the sealing ele-
ment
Contact with the technical
service for replacement
Follow point "7.Operation"
Follow chapter "7.Opera-
tion"
Check the assembly as
indicated in chapter "5.In-
stallation & Assembly". Lu-
bricate the frosted crystals
unions with sunflower oil.
Check connections as indi-
cated in chapter "5.Instal-
lation & Assembly".
tion"
Follow chapter "7.Opera-
tion"
Tel 34 937 700 877 Fax 34 937 702 362