Functional Overview
Rear Panel Overview
Power Connector

Figure 2-2: Rear panel

2-6
Figure 2–2 shows the rear panel of the waveform monitor. A brief discussion of
each connector follows the illustration.
The waveform monitor is designed to operate from a single-phase power source
with the neutral conductor at or near earth ground. Only the line conductor is
fused for over-current protection. A protective ground connection through the
grounding conductor in the power cord is essential for safe operation.
WARNING. Dangerous potentials are present on the Power circuit board. Do not
connect power to the waveform monitor if it is not enclosed in a prescribed
cabinet.
The waveform monitor operates from an AC line frequency of 50 or 60 Hz, over
the range of 90 to 240 VAC, without the need for configuration. Refer to
Appendix A: Specifications for additional information on power and environmen-
tal requirements.