LF Magnetic Field Meter
MG-300
HB4MG3000000

Advertisement

Table of Contents
loading

Summary of Contents for LATNEX MG-300

  • Page 1 LF Magnetic Field Meter MG-300 HB4MG3000000...
  • Page 3: Table Of Contents

    Table of Contents Introduction ..........1 Application ..........1 Features ............ 2 Identifying Parts ........2 Measurement Procedures ......3 Specifications ..........5 Battery replacement ........6 Safety Precaution ........6 End of life ..........7...
  • Page 4: Introduction

    MG-300 1 Introduction Safety precautions should be taken against electric appliances in places including medical institutions, schools and residential districts, where people usually stay for a long time, to prevent patients, babies and senior citizens from exposure to high electromagnetic waves.
  • Page 5: Features

    MG-300 3 Features Switch between the display of micro-Tesla and  mini-Gauss is available. Data hold (HOLD)、maximum (MAX) Hold function.  Range display(20,200,2000).  Low battery detector “ ”.  Over load display “OL”.  4 Identifying Parts 1. Sensor position.
  • Page 6: Measurement Procedures

    MG-300 5 Measurement Procedures Press to power on.  Select for µT unit or for mG unit.  Position the front (refer to Figure 1) of the meter to  measure the electromagnetic waves. Try to change the measurement angle or position for obtaining the highest reading value (please refer to Figure 2 to 6).
  • Page 7 MG-300 Figure 1 indicate the correct measurement direction of  the electromagnetic wave. Figure 2,3, 4, 5 and 6 indicate the incorrect  measurement direction of the electromagnetic wave EN-4...
  • Page 8: Specifications

    MG-300 6 Specifications Display: 3-1/2 digits LCD, maximum reading 1999.  Range: 200/2000 mG,20/200µT.  Resolution: 0.1/1 mG or 0.01/0.1 µT.  Frequency response: 30Hz to 300Hz.  Sensor: Single Axis  Accuracy: ±(2.5%±6dgt) at 50Hz/60Hz.  Over load: LCD display “OL”.
  • Page 9: Battery Replacement

    MG-300 7 Battery replacement WARNING If the symbol “ ” appears on the LCD, please replace the battery immediately Turn off the instrument.  Remove the battery cover  Replace the battery.  Install the battery cover.  8 Safety Precaution For cleaning the instrument use a soft dry cloth.
  • Page 10: End Of Life

    MG-300 9 End of life Caution: this symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal EN-7...
  • Page 12 EMR Shielding Solutions Inc. 399 Four Valley Drive, Unit 13, Vaughan, Ontario, L4K 5Y7, Canada E-mail: info@emrshieldingsolutions.com www.emrshieldingsolutions.com...