Heating (Activation/Conditioning) Without Stop; Custom: A Generic Pump For Customer Settings - SAES NEG POWER User Manual

Neg pump controller
Table of Contents

Advertisement

4.9.

Heating (Activation/Conditioning) without stop

Some applications require special getter pumps (HV) that work when the getter material is hot.
NEG POWER can be used to keep these pumps hot giving them the necessary power
continuously. To do this, t_hold =  is used (to set t_hold =  reduce t_hold down to 100 and
press down again). When t_hold =  the power supply is only stopped by pressing the stop
button.
4.10.

Custom: a Generic Pump for Customer Settings

NEG POWER software is preset to know the activation/conditioning parameter settings of the most
popular SAES getter pumps, like the GP SORB AC, the CapaciTorr, CapaciTorr HV and the
NEXTorr families (see the Summary Table of the NEG pumps controlled by the NEG POWER).
In the software there is also a Generic Pump named Custom.
If the used pump model isn't included in the list, Custom pump can be selected and settings of
suitable voltage, current, power and temperature must be set for the specific pump model.
The default settings for Custom pump are the following. These values can and shall be changed.
In case of standard 700 W module:
NAME
Tmax
Custom
500 °C
In case of low power 150 W module:
NAME
Tmax
Custom
500 °C
ATTENTION: SAES Getters is not responsible for damages due to improper use of the
power supply
M.HIST.0101.23 R2
Activation Mode
Vmax
Imax
50 V
4.5 A
Activation Mode
Vmax
Imax
20.0 V
4.5 A
Page 20 of 36
Pmax
Tmax
500 W
250 °C
Pmax
Tmax
100 W
200 °C
Conditioning Mode
Vmax
Imax
25 V
2.5 A
Conditioning Mode
Vmax
Imax
5.0 V
2.5 A
January, 26
Pmax
250 W
Pmax
15 W
th
2021