Carefully follow the user and maintenance instructions to ensure proper functioning and to extend the life of the device. ASPIRA Go (1) is equipped with a vacuum flow adjustment knob (4), vacuum gauge (3) and 1000 ml canister (9) with inflow safety device (10) to prevent fluids from entering the suction pump. This device interrupts the vacuum flow via a float-controlled valve found in the canister lid.
Page 5
• The power supply provided has been specifically designed for use with the Aspira Go series aspirators. Never use the power supply with other devices or for uses other than that specified in this manual and never use the Aspira Go series aspirators with other power supplies.
Flat batteries must be disposed of in special waste collection containers, or by contacting a suitable waste disposal centre. It is forbidden to use batteries other than those provided with the medical device. If necessary contact an authorised FLAEM service centre.
Instructions for suction on the patient: 3.1.Connect the accessories, referring to the "CONNECTION DIAGRAM" on the cover. ATTENTION: in addition to being antibacterial/antiviral, the hydrophobic filter also acts as secondary protection device and stops any fluids that the primary safety device in the canister lid cannot manage to stop.
Page 8
FILTER The hydrophobic/antiviral and antibacterial filter supplied with Flaem aspirators must be strictly replaced for each new patient, or if the said filter is saturated. Should the aspirator be used on the same patient, the filter for Single-patient use must be replaced within a maximum period of TWO MONTHS or if the filter is saturated.
Should anomalous situations occur, i.e. softening or cracking of the accessories, quickly terminate the procedure and substitute the used materials. Contact the authorised service centre and specify how the product was used. Note: Use only with original FLAEM accessories.
TECHNICAL SPECIFICATIONS Mod. P1611EM-20 / P1611EM-30 Voltage 12 V VEHICLE POWER OUTLET POWER CABLE COD. 16667 Internal battery supplied: 2500 mAh; 14,8 V; lithium-ion In the event of a fuse failure, re- 2500 mAh; 14,4 V; lithium-ion place with a 5A - 250V approved 2900 mAh;...
RF mobile or portable radio and telecommunications devices (mobile telephones or wireless connections) can interfere with the functioning of electro-medical devices. For further information visit our website www.flaemnuova.it. Flaem reserves the right to make technical and functional modifications to the product with no prior warning.
Dirty, blocked or damaged dealer or an authorised FLAEM pump service centre If, after checking the aforementioned conditions, the device still does not work properly, we recom- mend that you contact your local dealer or an authorised FLAEM service centre.