5) According to the
5.3 Theoretical value calibration
The theoretical value calibration refers to the weight calibration
operation through the input connection of sensor sensitivity and sensor range
value.
Theoretical value calibration requires three steps:
1) Set sensor sensitivity (for example, connect multiple sensors and
enter the average value of sensitivity)
2) Set the total sensor range (such as connecting multiple sensors,
input the total sensor range and)
3) Turn on the "theoretical value effective" switch
Key to complete weight calibration.
16