4. Calibration
4.1 Description
GM7702 indicator should proceed calibrated if the parameters can't meet users'
requirement. Calibration can fix zero point and gain voltage.
User can make record of calibration parameter value, which could be used in calibration
without weight.
4.2 Calibration Parameter
Calibration parameter
Unit
Decimal point
Division
Max. capacity
Remote calibration
ON/OFF
Calibration with
weight
Parameter
g;kg;t;lb
0;0.0;0.000.000;0.0000
1;2;5;10;50
≤division *100000
OFF or ON
empty scale calibration
weight value 1
- 22 -
GM7702 User Manual
Description
initial value: kg
initial value: 0.00
initial value: 1
initial value: 10000
Calibration through serial port
when set ON.
Press ENTER key and set present
status as zero point.
The weight value is base of the