Meter Does Not Turn Off (Problem 1); Meter Enters Wrong Function (Problem 2) - Test Equipment Depot 88 Service Manual

Automotive meter
Table of Contents

Advertisement

88
Service Manual
1. The Meter will not turn off when the rotary switch is in the OFF position.
OR
2. The Meter does not enter the function indicated by the rotary switch setting.
Both of these problems can be intermittent and can be caused by different malfunctions.

3-31. Meter Does Not Turn Off (Problem 1)

If the Meter does not turn off when the rotary switch is in the OFF position, the wiper
contacts are probably not making contact with the PTF sections. This may not allow the
Power Down Signal (PDS) line to be pulled low by VBT-. Check the wiper contacts for
good spring tension against the PTF surface. Make sure that the PTF surface and spring
contacts are clean. If they are not, clean them with isopropyl alcohol. To get to the wiper
contacts, perform the procedures under "Disassembling the Case" and "Removing and
Reinstalling the Circuit Assembly" previously described in Chapter 3.
This problem could be unrelated to the PTF, in which case a power supply component,
such as Q8, may be at fault. Refer to TROUBLESHOOTING THE POWER SUPPLY
earlier in Chapter 3.

3-32. Meter Enters Wrong Function (Problem 2)

If the Meter does not enter the function indicated by the rotary switch selection, the most
probable cause is an out-of-tolerance PTF resistor section. This can be checked by
performing the procedure under "Function Selection (Rotary Switch) Test" given earlier
in Chapter 3. If the display reading is outside the tolerances given, the PTF resistor
section is bad, and the pca must be replaced. (This assumes that the PTF wiper contacts
are good. If the wiper contacts are not making proper contact with the PTF sections, the
Meter can also select the wrong function. See "Meter Does Not Turn Off", above.
However, poor contact is more likely to cause an intermittent rather than a total failure.)
If the Meter powers up in ac mV and will not change functions, the problem is caused by
the inability of the a/d converter to measure, NOT BY THE PTF. The a/d converter must
do a ratio measurement of the PTF resistor in order to change functions. U4 or one of its
peripherals can cause this problem.
3-24

Advertisement

Table of Contents
loading

Table of Contents