Kanmed Operatherm 202 User Manual page 3

Table of Contents

Advertisement

Manual KanMed Operatherm
Key switch for temperatures <35ºC and >37ºC
Temperature selector switch
Alarm silence switch
Heating Pad failure alarm
Over temperature alarm
Control Unit failure
Earth potential equalisation connection point
2.3 EXPLANATION OF PICTURES ON THE HEATING PAD
1
Refer to users manual
2
Heating Pads OP-200-xxx may
only be used with Operatherm Control Unit OP-200/202
3
Unauthorised persons are not allowed to open the Control Unit or the Heating Pad Connector.
Any such interventions may endanger the patient safety.
4
The Heating Pad cable may not be altered in length or repaired in any other way.
5
See Users Manual section 10, concerning cleaning and maintenance of Heating Pads and
Control Units
6
Sharp objects must not penetrate the Heating Pad
7
Place the Heating Pad flat on the operating table with the text side towards the patient.
8
Gel Pads must be placed over the Heating Pad.
9
When used with children, ensure that the gel pad Gel Pads covers the whole Heating Pad
10
Ensure that the text side of the Heating Pad is towards the patient or ECG disturbance may
occur.
11
The Heating Pad should be stored using the Pad Hanger or rolled with a diameter not smaller
than 150mm. Don't bend or fold the Heating Pad so that sharp crease arises.
3
Page

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Operatherm 200

Table of Contents