USER INTERFACE SETTINGS (OS-MINI AND OS-MINI22)
Date & Time
Change the date and time for data logging purposes. The clock is reset when the
power is cycled unless a battery is fitted.
Output Processing
Averaging
Set the time, in seconds, over which the measured temperature is averaged. Note:
Period
averaging prevents the sensor from following rapid temperature changes. Minimum: 0
(no averaging). Maximum: 60.
Hold Mode
Peak
The sensor holds the maximum temperature steady for the Hold Period. After this, the
sensor responds normally. If the sensor detects a higher temperature, it holds this
temperature steady for the Hold Period.
Valley
Similar to Peak Hold mode except that the sensor outputs the minimum detected
temperature steady for the Hold Period.
Hold Period
Off
Disables hold processing.
Set the time, in seconds, for the sensor to hold the temperature as above.
Minimum: 0 (no hold processing). Maximum: 1200.
Data Logging
Sample
The time, in seconds, between samples. Minimum: 1. Maximum: 86,400 (1 day).
Period
Number of
The number of samples to collect before logging stops. Minimum: 0 (continuous
logging). Maximum: 86,400 (1 day of data if Sample Period = 1 second).
Samples
Enable
The sensor begins logging at the Date and Time specified. Logging can also be
Scheduled
started and stopped manually.
Start
The date and time for scheduled logging to start.
Date and
Time