TABLE OF CONTENTS 1. Introduction ....................2 2. Installation ....................2 2.1 Set Up Mounting Assembly ................ 4 2.2 Connect to METER ZENTRA Series Logger ..........5 3. Data Interpretation ................6 4. Troubleshooting ..................6 5. Customer Support ................7 Index ........................
Apogee IR radiometers purchased from METER come preconfigured by METER to work seamlessly with METER ZENTRA series data loggers. The information in this document explains how to integrate Apogee IR radiometers into the ZENTRA system.
Page 5
INSTALLATION Table 1 Installation (continued) Conduct System Check METER strongly recommends setting up and testing the system (sensors and data loggers) in the lab or office. Inspect and verify all components are intact. Visit the data logger product page for the most up-to-date software and firmware.
IR radiometers must be mounted such that their FOV encompasses the surface of interest. Each IR radiometer purchased from METER comes with the appropriate mounting bracket that can be mounted to either a horizontal or vertical post, depending on which set of holes is used.
Use ZENTRA Utility to set the measurement interval. Use ZENTRA Utility to configure communication settings for data transfer to ZENTRA Cloud. Sensor data can be downloaded from METER data loggers using either ZENTRA Utility or ZENTRA Cloud. Refer to the logger user manual for more information.
Customer Support for a replacement connector or splice kit. If a cable is damaged refer to the METER wire-splicing guide cable repair. It is recommended that Apogee IR radiometer sensors are returned for recalibration every 2 years. Visit Apogee repairs (apogeeinstruments.com/recalibration-and-repairs) or contact...
Phone: +49 89 12 66 52 0 Fax: +49 89 12 66 52 20 Website: metergroup.de If contacting METER by email, please include the following information: Name Email address Address Instrument serial number Phone Description of the problem...
INDEX INDEX cables 3, 5 ZENTRA series data loggers 2, 3, 5 connecting 5 ZENTRA software customer support 7 Cloud 2 Utility 2, 5, 6 ZSC 2 data 6 data logger. See ZENTRA series data loggers installation 2–3 mounting 3 preparation 3 tools needed 2 mounting 3...
Page 11
18482-00 8.31.2021 METER Group, Inc. USA 2365 NE Hopkins Court Pullman, WA 99163 T: +1.509.332.2756 F: +1.509.332.5158 E: info@metergroup.com W: metergroup.com METER Group AG Mettlacher Straße 8, 81379 München T: +49 89 1266520 F: +49 89 12665220 E: info.europe@metergroup.com W: metergroup.de...
Need help?
Do you have a question about the SI-400 Series and is the answer not in the manual?