Wrist Strap Tester 746
Instructions
European Version
March 2005
78-8136-2118-8

Advertisement

loading

Summary of Contents for 3M 746

  • Page 1 Wrist Strap Tester 746 Instructions European Version March 2005 78-8136-2118-8...
  • Page 3: Declaration Of Conformity

    3.0 Description Wrist straps are the primary method to minimize charge The 3M™ Wrist Strap Tester 746 operates as a battery generation on the human body. They must be able to powered tester, or additionally with an AC power supply drain this charge as rapidly as it is generated.
  • Page 4: Specifications

    Put the wrist band on with ground cord attached, and insert the ground cord plug into the wrist For removal, lift the tester 746 at the two top strap plug-in jack. corners with both hands.
  • Page 5: Calibration Procedures

    6.0 Calibration Procedures Wrist strap test circuit check: Connect the resistance substitution box to the 746 as The 3M™ Wrist Strap Tester 746 cannot be adjusted, shown in figure 2. but the following steps can be used to determine that the tester is operating within its specification.
  • Page 6 3M™ Wrist Strap Tester 746 Logbook 1 2 3 4 5 6 7 8 9 W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W...
  • Page 8 Warranty; Limited Remedy; Limited Liability. This product will be free from defects in material and manufacture for a period of one (1) year from the time of purchase. 3M MAKES NO OTHER WARRANTIES INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.