WARNING
NightBalance should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, NightBalance should be observed to verify
normal operation in the configuration in which it will be used.
NightBalance does not have essential performance.
Electromagnetic emissions
Emissions Test
RF Emissions
(CISPR 11)
RF Emissions
(CISPR 11)
Harmonic Emissions
(IEC 61000-3-2)
Voltage fluctuations / flicker
emissions
(IEC61000-3-3)
40
Compliance
Electromagnetic Environment - Guidance
Group 1
The NightBalance uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Class B
The NightBalance is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
Class A
that supplies buildings used for domestic purposes.
Complies
Regulatory Information