  • Page 1 Instructions for use NightBalance English (EU) Sleep position therapy...
  • Page 2: Table Of Contents

    Login and pair your Sensor Device Setting up the Docking Station Technical specifications......34 Symbols .............36 Understand the adaptation program How to use NightBalance daily ....16 Environmental conditions ...... 37 Placing the Chest Strap Regulatory information ......38 Placing the Sensor Device Data privacy &...
  • Page 3: Introduction

    Read all instructions for use before using NightBalance. Contact your physician if you have any questions as to whether NightBalance is right for you. You may also contact customer service if you have any questions about NightBalance, see page 43.
  • Page 4: Intended Use

    You should be able to switch sleeping positions independently. Contraindications for use • NightBalance should not be used by a patient using another medical aid that can be affected by mild vibrational stimuli - on the chest. • NightBalance is not recommended for the treatment of patients who are supposed to sleep in a supine position because of a medical condition (e.g.
  • Page 5: Warnings And Precautions

    Warnings & Precautions Warnings ⚠ NightBalance is only to be used for the purpose as specified in the ‘Intended Use’. Note all warnings and cautions throughout this user guide and on labels on the components. ⚠ NEVER wear or use the Sensor Device or Docking Station in a wet environment such as the shower or bath.
  • Page 6 Only use accessories, detachable parts and materials described in the Instructions for Use. Only use the Power Adapter provided by NightBalance (100-240 V, 50/60 Hz) included in your NightBalance pack. Use of other parts and/ or materials may damage the device. NightBalance cannot ensure continued safe use of a damaged device.
  • Page 7 CHARGE the Sensor Device at least once every other day for optimal performance. There is no harm in charging the Sensor Device every morning after use. ⚠ USE, and CHARGE NightBalance in temperatures between -5ºC and 40ºC (23ºF and 104ºF). STORE NightBalance in temperatures between -25ºC and 70ºC (-13ºF and 158ºF).
  • Page 8: Nightbalance Components

    NightBalance components Sensor Device Docking Station Navigation buttons Contact pins LED light Power port ON/OFF switch Display Front Front Back Back NightBalance components...
  • Page 9 Chest Strap Buckle Device pocket Power Adapter Cable length: 1.5m (4’11”) The Power Adapter and wall outlet as visualized in the instructions are an example only, they might look different. NightBalance components...
  • Page 10: Getting Started

    Getting started Step 1: Check packaging contents Upon receiving, make sure your packaging contains: Travel case Instructions for Use Chest Strap Docking Station Sensor Device Power Adapter If any of these items are missing upon receiving, contact your distributor or customer service (see page 43).
  • Page 11: Install The App And Register

    Portal to register your Sensor Device. Go to the online Portal at eu.mynightbalance.com and follow the instructions on the Portal. Download the Philips NightBalance 07:42 App on your smartphone from the Play Store (for Android) or App Store (for iPhone).
  • Page 12: Login And Pair Your Sensor Device

    Step 3: Login and pair your Sensor Device Login to the Philips NightBalance app after activating your account. Follow the instructions in the app to complete your account. Once indicated by the app, activate the pairing mode by switching the Sensor Device ON while pressing one of the navigation buttons.
  • Page 13: Setting Up The Docking Station

    Step 4: Setting up the Docking Station Insert the Power Adapter cable into Connect the Power Adapter into a the back of the Docking Station. standard wall outlet (100-240V AC). Getting started...
  • Page 14 The display and LED of the Sensor Docking Station with the buttons on Device should become active. This top and display towards you. indicates the right placement. NightBalance is now ready for use. On the Docking Station, the Sensor Device charges itself automatically. Getting started...
  • Page 15 Sensor Device’s vibrations during 2) Build-up phase sleep. You need to use NightBalance for NightBalance starts to deliver vibrations nine nights before the device will provide during night 3 to 9. The vibrations will full position therapy.
  • Page 16: How To Use Nightbalance Daily

    How to use NightBalance daily Step 1: Placing the Chest Strap Wrap the Chest Strap around your Hook the buckle through the opening chest In case the Chest Strap is too short, adjust the tightness (see next page). How to use NightBalance daily...
  • Page 17 Make sure the Device pocket is placed adjusting the velcro position of the in the middle of your chest. Chest Strap. The Chest Strap should fit tight enough to stay in place and feel comfortable to wear at the same time. How to use NightBalance daily...
  • Page 18: Placing The Sensor Device

    Step 2: Placing the Sensor Device Power on the Sensor Device by sliding Confirm that the white LED light on the ON/OFF switch to the ON position. the Sensor Device starts blinking. This indicates it is switched on. How to use NightBalance daily...
  • Page 19 Check that the Device pocket is placed facing away from your body. on the middle of your chest. Now you can go to sleep. Display is not facing Buttons are not on away from your body How to use NightBalance daily...
  • Page 20: Sleeping With The Sensor Device

    Step 3: Sleeping with the Sensor Device Fall asleep in any position you desire. Once the Sensor Device vibrates, turn to your side. NightBalance allows you to fall asleep in any position. 15 minutes after NightBalance regulates the intensity of switching the Sensor Device ON, the the vibrations based on your sleeping therapy starts.
  • Page 21 Device pocket of the Chest Strap vertical for a second. In pause mode, the Sensor Device will not give vibra- tions for 5 minutes. Activating Activating pause mode pause mode How to use NightBalance daily...
  • Page 22: Removing The Sensor Device

    View last night’s sleep data on the Strap. Sensor Device’s display if desired. Take the Sensor Device out of the Sleep data from short naps (<2 hours) Chest Strap. is excluded from any data calculations. How to use NightBalance daily...
  • Page 23 Sensor Device the night the vibrations of the last night before Sensor Device Press the navigation buttons on top of the Sensor Device to switch from one screen to the next. How to use NightBalance daily...
  • Page 24 Power off the Sensor Device by Confirm that the white LED light on sliding the ON/OFF switch to the OFF the Sensor Device stops blinking. This position. indicates it is switched OFF. How to use NightBalance daily...
  • Page 25: Charging The Sensor Device

    Docking Station, the Sensor Device buttons on top and display towards display will become active and the you. LED will start to blink. The display of the Sensor Device shows a battery icon when charging takes place. How to use NightBalance daily...
  • Page 26: View Therapy Data

    OFF again. Synchronizing should start Sensor Device ON or OFF during daily automatically. use when the Philips NightBalance app has been downloaded to your smart- Synchronization with the smartphone phone. For synchronization progress on a daily basis ensures your therapy and information, see the Philips data is up to date.
  • Page 27: View Therapy Data In The App And Portal

    Step 2: View therapy data in the app and Portal Viewing therapy data is possible on the Sensor Device, smartphone (Philips Last night Last week NightBalance app) and the online Portal. The following data views are available: Sensor Device: Last night...
  • Page 28 Portal once you grant them access. See the ‘Profile’ section in the Philips NightBalance app or Portal. Logout from the online Portal after use. View therapy data...
  • Page 29: Care And Maintenance

    Chest Strap. • It is the user’s responsibility to keep NightBalance equipment dust free. Before cleaning, turn the Sensor Device OFF and unplug the Power Adapter from the Docking Station and the wall outlet. Clean the Sensor Device and Docking Station with a dampened cloth with water or mild cleaning solution (e.g.
  • Page 30: Disposal

    The Sensor Device and Docking Station do not contain any parts that can be serviced. • After the service life (equivalent to in-use life on page 34 and 35) of NightBalance components has expired, replace the component with a new one. Care and maintenance...
  • Page 31: Troubleshooting

    Sensor Device should be fully charged. See page 13, 14, and 25. The Sensor Device shows the An error occured in Do not use NightBalance. following screen: the Sensor Device Contact your distributor, or customer service (see page 43), and mention the error number that your Sensor Device shows.
  • Page 32 Problem Why it happened What to do I cannot pair the Sensor Device Bluetooth on your Enable Bluetooth on your with the Philips NightBalance smartphone is smartphone. app and the Sensor Device switched OFF shows the following screen: The Philips...
  • Page 33 Problem Why it happened What to do My sleep data is not visible in The Sensor Synchronize the Sensor the Philips NightBalance app or Device has not Device with your smartphone. Portal. yet synchronized Follow the instructions on with the Philips page 26.
  • Page 34: Technical Specifications

    (lifetime: 3 years) Frequency Bluetooth: characteristics: version 4.2, 5.0 Travel Case Band support: 2.4GHz Accessory of NightBalance Effective radiated Dimensions: 20.0 x 17.0 x 6.8 cm power: +6 dBm (7.9 x 6.7 x 2.7 inch) Weight: 250 gr (8.8 oz)
  • Page 35 149.0 x 7.0 x 0.3 cm (58.7 x 2.8 x 0.1 inch) L2US01US Weight: 62 gr (2.2 oz) L2EU01NL Materials: Polyamide, Polyester, NightBalance L2EU01FR Lycra, Polyurethane, L2EU01DE Silicone L2EU01UK In-use life: 1 year IEC 60601-1 Classifications: This device is Class II Equipment, Applied Part Type BF, Continuous Operation.
  • Page 36: Symbols

    Symbols CE mark Refer to Prescription instruction Type BF medical manual/booklet Manufacturer equipment Waste Electrical and Electronical Serial Number Caution Equipment General warning Batch code Do not bleach sign Ingress Reference code IP22 Do not iron protection Keep dry Do not dry clean Polarity Non-ionizing Do not tumble...
  • Page 37: Environmental Conditions

    Environmental conditions Storage conditions Operating conditions NightBalance NightBalance components can components can be safely stored be safely used at temperatures at temperatures between -25 C (-13 between -5 C (23 and 70 C (158 F), a and 40 C (104...
  • Page 38: Regulatory Information

    Electromagnetic compatibility (EMC) NightBalance is intended for use in the electromagnetic environment specified in the coming sections. The customer or the user of NightBalance should assure that it is used in such an environment. NightBalance needs special precautions regarding EMC according to the EMC information provided in this chapter.
  • Page 39 WARNING: Portable and mobile RF communications equipment including antennas can affect NightBalance’s operations. DO NOT use portable and mobile RF com- munications equipment closer than 30 cm (12 inches) to any part of NightBalance, including cables specified by the manufacturer. Other equipment, even if that other equipment complies with CISPR emission requirements, may interfere with NightBalance.
  • Page 40 RF Emissions Class B The NightBalance is suitable for use in CISPR 11 all establishments, including domestic establishments and those directly connected to the public low-voltage power...
  • Page 41 Electromagnetic immunity Immunity test IEC 60601 Test Compliance Electromagnetic Environment - Level Level Guidance Electrostatic ±8 kV contact ±8 kV contact Floors should be wood, concrete discharge (ESD) or ceramic tile. If floors are IEC 61000-4-2 covered with synthetic material, ±2 kV, ±4 kV, ±2 kV, ±4 kV, the relative humidity should be at...
  • Page 42 Voltage <5 % U <5 % U dips, short (>95 % dip in U (>95 % dip in U interruptions for 0,5 cycle for 0,5 cycle and voltage variations on 40 % U 40 % U power supply (60 % dip in U (60 % dip in U IEC61000-4-11 for 5 cycles...
  • Page 43: Warranty

    Contact NightBalance B.V. provides a two-year Manufactured by NightBalance B.V. warranty on the Sensor Device and Docking Station. NightBalance or the reseller will Benoordenhoutseweg 46-13 repair or replace the device(s) due to faulty 2596 BC, The Hague materials or operation during the warranty The Netherlands period.
  • Page 44 Directive 93/42/EEC Concerning medical devices Instructions for use for NightBalance sleep position therapy Document number: DOC-1060 rev1.0 (EU English) © 2018 NightBalance B.V. All rights reserved.

Table of Contents