Macurco TX-6-AM User Instructions page 19

Ammonia detector, controller and transducer
Hide thumbs Also See for TX-6-AM:
Table of Contents

Advertisement

Available languages

Available languages

The display will flash during the test, or in the case of the gas test, the gas level will alternate with gtS. Once the test is complete, the
display will return to steady display. To exit the test menu, press the Next button until "End" is displayed, then press Enter to return to
normal mode.
Ammonia Gas Test
The following steps must be performed when conducting a calibration or calibration verification test (bump test) to ensure proper
performance of the monitor. Failure to do so may adversely affect product performance and result in sickness or death.
When performing a calibration or calibration verification test (bump test) only use certified calibration gas at the
required concentration level. Do not calibrate with expired calibration gas.
If the instrument cannot be calibrated, do not use until the reason can be determined and corrected.
Do not cover or obstruct display or visual alarm cover.
Ensure sensor inlets are unobstructed and is free of debris
General
The TX-6-AM can be bump-tested or calibrated with the AM-FCK with Ammonia gas, regulator and test hood, available through your
local representative or from Macurco.
Contents of the FCK
AM-FCK: (1) Gas Cylinder, 25 ppm Ammonia gas in air
Gas regulator with about two feet of plastic tubing
Gas test hood
FCK Information
Several detectors can be calibrated with one FCK. The only limitation is the amount of gas in the cylinder and the flow of the regulator.
The 34 liter cylinder for example with a 0.5LPM regulator has approximately 68 minutes of continuous calibration run time.
Replacement cylinders are available. The gas cylinder should be replaced when the pressure gauge on the regulator shows 25-psi or
less.
Note: For optimum test results it is suggested that the unit be in clean air (green light on) and be in a low ambient air flow
Gas Testing
Testing the Fan Relay –
1.
Remove the Philips screw on the front of the TX-6-AM. Remove the front cover.
2.
Open the FCK. Connect the 25 ppm gas cylinder to the regulator.
!
WARNING
19

Advertisement

Table of Contents
loading

Table of Contents