Specifications - Omron M1 Basic Instruction Manual

Automatic upper arm blood pressure monitor
Hide thumbs Also See for M1 Basic:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 4

6. Specifications

Product description
Product category
Model (code)
Cuff pressure range
Blood pressure measurement range 20 to 280 mmHg
Accuracy
Inflation
Measurement method
IP classification
Rating
Power source
Battery life
Durable period (Service life)
Operating conditions
Storage / Transport conditions
Contents
Protection against electric shock
Weight
Dimensions (approximately value) Monitor: 112 mm (W) × 82 mm (H) × 140 mm (L)
Cuff circumference applicable to
the monitor
Memory
Note
• These specifications are subject to change without notice.
• This monitor is clinically investigated according to the requirements of ISO 81060-2:2013 (excluding pregnant and pre-eclampsia patients). In the
clinical validation study, K5 was used on
85 subjects for determination of diastolic blood pressure.
• IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are
protected against solid foreign objects of 12.5 mm diameter and greater such as a finger. The optional AC adapter HHP-CM01 is protected
against vertically falling water drops which may cause issues during a normal operation. The optional AC adapter HHP-BFH01 is protected
against oblique falling water drops which may cause issues during a normal operation.
• Operating mode is classification in accordance with IEC 60601-1.
Automatic Upper Arm Blood Pressure Monitor
Electronic Sphygmomanometers
M1 Basic (HEM-7121J-AF)
0 to 299 mmHg
Pressure: ±3 mmHg / Pulse: ±5% of display reading
Automatic by electric pump
Oscillometric method
Monitor: IP20 / Optional AC adapter: IP21 (HHP-CM01) or IP22 (HHP-BFH01)
DC6 V 4.0 W
4 "AA" batteries 1.5 V or optional AC adapter
(INPUT AC 100 - 240 V 50 - 60 Hz 0.12 - 0.065 A)
Approximately 1000 measurements (using new alkaline batteries)
Monitor: 5 years / Cuff: 5 years / Optional AC adapter: 5 years
+10 to +40 °C / 15 to 90% RH (non-condensing) / 800 to 1060 hPa
-20 to +60 °C / 10 to 90% RH (non-condensing)
Monitor, arm cuff (HEM-CR24), 4 "AA" batteries, Instruction Manual
Internally powered ME equipment (when using only batteries)
Class II ME equipment (optional AC adapter)
Monitor: approximately 250 g (not including batteries) / Arm cuff: approximately 130 g
Arm cuff: 145 mm × 466 mm (air tube: 600 mm)
170 to 420 mm (included arm cuff: (220 to 320 mm))
Last measurement
Display
LCD digital display
Pulse measurement range 40 to 180 beats / min.
Deflation
Automatic pressure release valve
Operating mode
Continuous operation
Applied part
Type BF (arm cuff )
and
EN
EN8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hem-7121j-af

Table of Contents