View Calibration Values (F18); Key-In Zero Calibration Value (F19); Key-In Span Calibration Value (F20); Calibration Value Table - Salter Brecknell 400 Series Setup & Operation Manual

Digital indicator
Table of Contents

Advertisement

6.4

VIEW CALIBRATION VALUES (F18)

Note: The values displayed in this procedure are valid only after a successful calibration
has been performed using F16 and F17.
1. While in the Setup mode, scroll to "F 18", then scroll down once using the #2 key to
enter View calibration menu.
2. The display will momentarily show "CAL 0" followed by a value. This value is the zero
calibration value and should be recorded in the table below. Press any key to
continue.
3. The display will momentarily show "CAL 1" followed by another value. This value is the
span calibration value and should also be recorded in the table below. Press any key
to return to upper level (F18).
INDICATOR
S/N:
6.5

KEY-IN ZERO CALIBRATION VALUE (F19)

Note: This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid zero calibration value, obtained from a successful F16
calibration procedure, must be used.
1. While in the Setup mode, scroll to "F 19", then scroll down once using the #2 key. The
display will momentarily show "ET C 0", followed by a value of zero
2. Use the front panel keys to key-in the actual zero calibration value. If you make a
mistake, press the ZERO key to clear your entry and start over.
3. After entering the exact value, press the NET/GROSS key to save the value. The
display will show "E C 0" momentarily, then revert back up to F19.
6.6

KEY-IN SPAN CALIBRATION VALUE (F20)

Note: This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid span calibration value, obtained from a successful F17
calibration procedure, must be used.
1. While in the Setup mode, scroll to "F 20", then scroll down once using the #2 key. The
display will momentarily show "ET C1", followed by value of zero.
2. Use the front panel keys to key-in the actual span calibration value. If you make a
mistake, press the ZERO key to clear your entry and start over.
3. After setting the exact value, press the NET/GROSS key to save the value.
4. If the entered value is greater than zero, the display will show "E C 1" momentarily,
then revert back up to F20. If a value of zero is entered, the indicator will briefly show
"Err 5", then revert back to the screen described above in Step # 2.
ZERO CALIBRATION VALUE
Table 6-1: Calibration Value Table
Page 6 - 2
SPAN CALIBRATION VALUE