Bosch Vivatmo website at www.vivatmo.com. Have your serial number of your handheld or your basestation available when you contact the service center. You find the twelve digit code on your Vivatmo pro handheld, next to the symbol at the bottom of the device.
Initial information Safety information WARNING The following conditions can cause malfunction or disturbance to the Vivatmo pro and the dispos- able mouthpiece: • High humidity, extreme temperature (see “9.1 Device data”). • Direct sun radiation or high exposure to dust or volatile substances, e.g., disinfectants or nail polish remover.
Page 11
• Vivatmo pro shall not be used in unprotected networks or work environments. Data protection is under the responsibility of the organization using Vivatmo pro and can be supported by the use of credentials on your network. When connecting the basestation to the local area network consider that data is transmitted unencryptedly with the risk of unprotected access to patient data.
(other than asthma). Vivatmo pro should not be used in critical care, emergency care or in anesthesiology. The following conditions can infl uence correct measurement of results and shall be avoided: smoking or tobacco consumption for at least 1 hour before the mea- surement, eating or drinking at least 1 hour before the measurement, especially nitrate rich food (e. g., spinach), strenuous...
Manage patients, measurements and orders C O N T I N U E C O N T I N U E Welcome to Vivatmo pro, Jouko Kinnunen Charging status handheld battery Handheld connected M E A S U R E...
System elements Handheld elements On the handheld, you see the following operating elements: Disposable mouthpiece: Interface to breathe into the Vivatmo pro handheld. Display: Shows current and stored measuring data and device messages. ON/OFF button: Activate and deactivate the device.
The measurement procedure is guided by an animation on the display of the basestation or the LED on the back of the hand- held. Synchronize measurement results of stand-alone measurements with the Vivatmo pro handheld by placing the switched on Vivatmo pro handheld in the cradle of the Vivatmo pro basestation.
Page 19
Ambient NO measurement in progress. The memory is cleared. The Vivatmo pro handheld is connected to the Vivatmo pro basestation. The icon disappears when the Vivatmo pro handheld is used as a stand-alone device without the Vivatmo pro basestation. The audio sound is activated. The icon appears when audio sound is activated.
Installing the handheld To install the handheld, insert the rechargeable battery. 1. Open the battery compartment on the back of the Vivatmo pro handheld. 2. Hold the handling tab with two fi ngers and insert the rechargeable battery into the battery compartment. Mind the triangle marks on the battery and inside the battery compartment.
User accounts protect your patient data by access control to the system. When the user management is activated, the access is protected by the user name and a numeric passcode. You can use Vivatmo pro also when the user management is deactivat- ed.
Avoid connection of the Vivatmo pro basestation to a multiple socket-outlet to prevent interference by other devices. 1. Connect the jackplug from the power supply unit into the power adapter port on the rear of the Vivatmo pro basestation. 2. Plug the suitable adapter for your country firmly on the power supply unit.
Installation Software setup When you start your Vivatmo pro basestation for the fi rst time, you will be guided through a setup procedure of 5 simple steps, which helps you to enter all the required settings. 1. Switch on the basestation, by pressing the ON/OFF button.
The User administration screen opens. C O N T I N U E You need at least one administrator account to use Vivatmo pro. Remember this access code carefully. The reset of the administrator account can only be done by service personnel.
B0:72:BF:0F:E3:66 2. On the screen, select the FINISH button.The setup procedure is completed. Login When user accounts are enabled, Vivatmo pro demands authentication on the system. You have to login before you can start using the system. Instructions for use...
Page 26
The Home screen opens. • Select the MEASURE button to start a measurement, see “4 Measure”. Welcome to Vivatmo pro, Jouko Kinnunen • Select the MANAGE button to manage the data, see “5 Manage”. M E A S U R E...
Measure WARNING If the Vivatmo pro device has been stored in warm and humid environment or not been in use for a longer time, reconditioning may be required and starts automatically during first measurement. Therefore per- form a test measurement every day before using the device with the patient, see “7.3 Reconditioning”.
The measurement procedure is guided by the handheld display and the LED on the upper side. Synchronize measurement results of stand-alone measurements with the Vivatmo pro handheld by placing the switched on Vivatmo pro handheld in the cradle of the Vivatmo pro basestation.
Possible interpretation of results according ATS Guideline 2011 and ATS guideline 2021, see chapter “11.3 Literature”. Displaying stored measured values Your Vivatmo pro handheld automatically stores up to 1,000 measured FeNO values, identified by time, date and measuring ID. The measured values are stored in descending order starting with the latest.
The access to settings depends on the login as professional or administrator, see “3.2.1 User concept”. 1. On the Home screen, select the Settings menu. G E N E R A L Welcome to Vivatmo pro, Jouko Kinnunen Welcome to Vivatmo pro, Jouko Kinnunen...
Quality control (QC) External QC tests are used to control the measurements of the Vivatmo pro handheld in comparison to reference concentra- tions of nitric oxide (NO). The system documents the tester who performs the QC tests by the user login.
Page 56
You see the list of active handhelds. If no handheld is active, you see the message: “No handhelds available”. 6. Compare the number of the Vivatmo pro handheld in the list with the serial number beneath the symbol on the handle of the handheld.
6.5.6 Network confi guration Vivatmo pro can be connected to a local area network to receive measurement orders, send back measurement results, export patient related measurement data, or use a network printer. Vivatmo pro supports Ethernet and Wireless via WLAN connection.
Page 59
Settings Supported confi guration of an ethernet connection 1. Plug in the ethernet cable in the ethernet port on the rear of the Vivatmo pro Network Configuration basestation. Select Your Networksetup 2. On the Network screen, select Confi gure Network.
6.5.7 Order/result interface Vivatmo pro can receive measurement orders from and send results to an electronic patient record system (EPR), if the net- work is confi gured. The interface standards GDT (German Data Transfer) and HL7 (Health Level 7) are supported.
Vivasuite is the digital Vivatmo ecosystem allowing use of services such as device management and remote software up- date. Vivasuite runs in the Bosch managed Cloud and applies highest standards regarding IT security and data privacy. The Vivasuite service is not available in all countries. Ask your local distributor.
Page 67
1. Vivatmo pro: On the Network/Interfaces screen, select Vivasuite Vivasuite Configuration Confi guration. Language 2. Vivatmo pro: Select Disconnect. You are asked to confi rm before device gets Connection Status Connected disconnected. D I S C O N N E C T...
Maintenance and trouble shooting The Vivatmo pro is easy to maintain. Clean the device in accordance with the hygienic standards of your organization. Cleaning and Disinfecting 1. Turn off the device. Wash hands thoroughly with soap and water. 2. Wipe the entire Vivatmo pro with alcohol disinfectant wipes with maximum 30 % alco- hol, e.g., Caviwipes1 (Metrex), Mikrobac Tissues (Bode Chemie GmbH) or mikrozid uni-...
Reconditioning Vivatmo pro performs an automatic stability check with the fi rst measurement of the day: If the Vivatmo pro device has been stored in a warm and humid environment or has not been in use for a longer time, the wrench symbol may appear on the basestation and the handheld.
Remote Software Update When the Vivatmo pro is connected to Vivasuite (see “6.5.9 Vivasuite confi guration”) new software updates will automatically be available on the device. The device administrator always has full control over new software updates. A notifi cation informs about the availablitiy of a new software update and the administrator can perform the software update whenever it fi ts the schedule.
Vivatmo pro handheld, see “7.2 Status information hand- possible. held”. Number of permissible measurement Connect a new Vivatmo pro handheld and dispose of the old trials reached. Device does not perform one, see chapter “8.1 Disposing the Vivatmo pro device and the any measurements after maximum battery”.
Page 77
Solution Humidity in device too high. Check ambient conditions and take a new mouthpiece. Ambient temperature and air pressure Use the Vivatmo pro in a location with permissible ambient (altitude) is outside the specified conditions. range. When transporting the device from one place to another, a stabilization period of at least 20 minutes should be observed before use.
Disposal Disposal Disposing the Vivatmo pro device and the battery Delete data on your device prior to disposal or service returns ,see “6.5.1 Factory Reset” and “6.5.5 Release the handheld”. For the purpose of disposal, please separate this device from other waste to prevent possible harm to the environment or human health from uncontrolled waste disposal.
Linearity Three Vivatmo pro devices were tested on linearity as part of a study on accuracy. A range of contrived FeNO concentrations (in total, nine NO concentrations and a negative control) from 3 ppb to 330 ppb were tested to determine linearity. The data showed that the test is linear within the claimed range of 5–300 ppb NO, with a slope of 1.0 ±...
Page 82
Within-device precision = estimate of variation between test runs and days The precision and repeatability for Vivatmo pro is claimed to be a standard deviation of ≤ 5 ppb for readings ≤ 50ppb, and a CV of ≤ 10 % for readings > 50 ppb and ≤ 160ppb, and a CV of ≤ 15 % for readings > 160 ppb. The precision/repeatability testing performed confirmed that the precision/repeatability performance of Vivatmo pro fell within these specifications.
Analytical Accuracy Three Vivatmo pro devices were tested at nine NO concentrations between 3 and 330 ppb NO with two test runs and two replicates per device per run. The data showed that mean Vivatmo pro results were within 5 ppb for NO concentrations ≤...
Page 84
Two clinical precision studies were performed with a total of 143 subjects (65 children and 78 adults). In the first study Participants were asked to obtain two Vivatmo pro measurements in two visits each for a total of N = 88 Vivatmo pro evalua- tions with 2 measurements.
Page 85
For those with elevated initial FeNO levels > 30 ppb for adults and > 25 ppb for children (total n = 95),there was a fall in mean FeNO measured by Vivatmo pro in patients with elevated FeNO levels for combined adult and pediatric treatment populations.
Direct current The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Bosch Healthcare Solutions GmbH is under license. Warranty The statutory provisions on warranty rights in consumer goods sales in the country of purchase shall apply.
Page 89
Portable RF communication devices (including peripherals such as antenna cables and external antennas) are not to be used closer than 30 cm next to any part of the Vivatmo pro system. The Vivatmo pro is intended for use in professional healthcare facility environment. The customer or the user of Vivatmo pro should assure that it is used in such an environment.
The use of accessories, transducers and cables other than those specified may result in increased emissions or decreased immunity performance of the equipment or system. The following list shows cables, transducers, and other applicable accessories with which Bosch Healthcare Solutions claims EMC compliance.