KMG-Lite-BT. The KMG-Lite-BT is equipped with a wireless module and therefore deploya- ble with the free KMG-Lite App. Contents that are not related to the KMG-Li- te app and the wireless module also apply to the KMG-Lite-BT. KMG-Lite can be retrofitted with a wireless module.
Page 5
Identification Display Meaning Note Note Offers helpful tips and information Objective Denotes the objectives that is to be achie- ved via the steps descripted Step Carry out steps Cross reference Refers to more extensive or detail Display Shows the display output...
Safety information This chapter gives general rules which must be followed and observed if the product is to be handled safely. Â Product safety/preserving warranty claims ► Operate the instrument only within the parameters specified in the chap- ter “Technical Data”. (Chapter. 12, S. 54) ►...
Page 7
 Ensure correct disposal ► Take used batteries to the designated collection points. ► Send the product to GTE Industrieelektronik GmbH at the end of its service life. L We will take care of its proper and ecological disposal..
Intended Use KMG-Lite and KMG-Lite-BT are compact handheld measurement instruments for determining dynamic closing forces of power-operated doors and gates, according to EN 12453, EN 16005 and ASR A1.7. ► Only use the product for those applications which it was designed for.
Product description This chapter describes the structure of the product and its display and control elements. Point of force application Display Control button Battery plug...
Page 10
Numeric measurement result Indicator Indicator Indicator...
Commissioning Follow the steps described to commission the device correctly. . Â Removing the protective film on the display: ► Peel the protective film off carefully. Â Inserting batteries: ► Remove battery plug at the end of the hand- le tube by twisting it counter clockwise and remove it.
Page 12
► Insert three AA - type batteries with 1.5V each. ► Mind the polarity! ► Close battery plug: Press the batteries with the battery plug against the spring resistance inside the handle and lock it by twisting it clockwise. Battery positive terminal ahead! 3 x batteries AA 1,5 V...
Page 13
 Turn on the KMG-Lite: ► Press the control button once The device performs a self-test during start-up.. Startup test The device is ready for measurement after the self-test (approx. 3 seconds).. Ready for measurement...
Measuring This chapter describes the steps required to measure with the product. Infor- mation on the measurement process in conjunction with the KMG-Lite app can be found on page 33. Â Start measurement: ► Start device (S.13) ► Wait for measure attendance.
Page 15
L Simultaneously with the measurement, the normative evaluation of the temporal force curve starts. Measurement/Evaluation  Display of the measurement results L After the measurement and evaluation are finished, the instrument auto- matically displays the dynamic peak force (F ) that has occurred wit- hin measurement.
Page 16
If the dynamic time (t ) exceeded the normative 0.75 seconds, the LED within the second display segment flashes. Measurement result: = 326 N > 0,75 s If the force value at the end of the 5s measurement period (F ) exceeded 25 N, the LED within the third display segment flashes.
Page 17
If the force value at the end of the 5s measurement period (F ) exceeded 80 N, the LED within the third display segment flashes. Measurement result: = 326 N > 80 N If the force value at the end of the 5s measurement period (F ) exceeded 25 N and the dynamic time (t ) exceeded 0.75 s, the LEDs within the...
Page 18
If the force value at the end of the 5s measurement period (F ) exceeded 80 N and the dynamic time (t ) exceeded 0.75 s, the LEDs within the second, third and fourth display segment flash. Measurement result: = 326 N >...
Page 19
 Display of the value F ► After measurement press the control button twice. The display now shows the value of the final force in Newton (N). = 0 N L The LED within the third display segment is lit for orientation. L By pressing the control button again, the display returns to the dynamic peak force with the flashing segments for indication of t and F...
Page 20
L Regardless of the current display state the instrument is always ready to start a new measurement whenever the required trigger force > 20 N is detected. Â Switch off the device ► Keep the control button pressed for more than two seconds. L After five minutes of inactivity the device automatically switches off.
Warning- and error signals The device reports special system states via short view on the 4-digit display. Exceedance of measurement range If the measured dynamic peak force exceeds the specified upper limit of 1600 N, the display shows a flashing “1600”. The measurement results for and F are not affected.
Page 22
Low battery warning If the battery voltage falls below a critical value, the display shows the mes- sage ”Batt“. L By pressing the control button you can still continue your measurements. ► Change the batteries as soon as possible. Error signal „Load rejection“ If the device senses a zero force offset on startup the display shows the mes- sage “Err“.
Page 23
cified temperature/humidity range or if the force application plate got stuck. ► Make sure the device is unloaded while starting up. ► Make sure that all environmental conditions are within the specified range. ► Make sure that the force application plate is neither stuck or soiled. (chap.
Page 24
► If this error message should occurr although the instrument is used within the specified temperature range, please send it to GTE for calibration. ( chap.10, page 51) Error signal “Eich“: calibrationtable missing If the instrument’s display shows the message “Eich“ after startup, the system is unable to find its calibration reference.
L Please note that the KMG-Lite app can only be used in conjunction with a KMG-Lite retrofitted with wireless modules or the KMG-Lite-BT. ► If your KMG-Lite device does not have a wireless function, contact...
Page 26
Updates to the KMG-Lite app The KMG-Lite app is updated regularly. L Updates to the KMG-Lite app are displayed under the menu item „Set- tings“ and can be switched on by the user on request by clicking on the respective new function.
Page 27
 Install the app ► Download the KMG-Lite app from the App Store for iOS and Google Play Store for Android devices. L Alternatively, this QR code can be used: Link: www.app.kmg-lite.com ► Follow the instructions for installation!
Page 28
 Connect to KMG-Lite-BT ► Switch on wireless connection on your smartphone or tablet. L The KMG-Lite-BT only connects to the smart phone or tablet during mea- surement. (see also „Select device“ on page 32)  Start the KMG-Lite app...
Page 29
Symbol Explanation Symbol Name Explanation Open user form Client Select customer Close Close measurement Save Save measurement Protocol Create a protocol Mail Send protocol by mail Share Sharing protocol via different channels Retrieving Call up the calibration date of the device Delete Delete information Not rated...
Page 30
 Enter the contact details of the examiner ► Open the „menu box“ in the upper left corner. ► Use the button „Pen“ to open the input field „Name of Tester“ for data input. ► Using the „Select image“ button, you can upload a photo or a gra- phic with your logo from the memory of your device.
Page 31
 Customer management/Create customer ► Open the „Menu field“. ► Select „Customer Management“ to see a list of all created customers. ► Use the „Create customers“ button to create a new customer. To do this, enter the name of the customer and then click the „Create“ button. L The customer name can be deleted at any time by swiping from the right to the left.
Page 32
 Select device ► Open the „Menu field“. ► Select „Settings“. L Under „Measurement Device“, the last connected device is displayed. ► Press the „Update“ button to find and save a new device.  Reset data ► Open the „Menu field“. ►...
Page 33
L Select this option for documentation according to ASR 1.7. ► Press the „Measure“ button on the app‘s home page. Alternatively via „Mesurement“ in the „Menu field“. ► Turn on the KMG-Lite-BT. ► Press the button „Add measure“. ► The KMG-Lite-BT connects to the KMG-Lite app.
Page 34
L If the connection is successful, the dot flashes in the 1st segment of the L KMG-Lite-BT. L The KMG-Lite app displays the note „Ready to measure“. The measure- ment can take place. L If there is a connection problem between KMG-Lite-BT and the KMG-Lite app, connect the KMG-Lite-BT as a new device.
Page 35
L three individual measurements can be performed per measurement. ► Start another measurement within a measurement using the „Add mea- sure“ button. L In the „Details“ tab you can store information about the measurement. They are shown in the protocol. ►...
Page 36
 Use the measurement assistant ► Open the measurement assistant via the menu. ► Choose between a standard measurement or DHF measurement. L The standard measurement is suitable for the documentation of 3 indivi- dual measurements of a test point according to EN 12453. L The DHF measurement is aimed at customers from U.K.
Page 37
► Follow the further steps of the app and add measuring points as spe- cified. L If there are several measurements within one measuring point, the app automatically determines an average.
Page 38
Display of the measurements – The green comfort shade shows the normal range of the force F and the duration T The comfort shading is only displayed at the first measuring point. The comfort shading is hidden from the second measuring point. –...
Page 39
Edit measuring points  Rename or delete measuring point ► Open the measurement and select the „pen“ icon of the measurement point; the „Edit measurement point“ window opens. ► „Save“ will take over the new name. ► „Cancel“ will leave the respective graphics window and the changes will be discarded.
Page 40
 Assess the measurement result ► Use the „-“ switch to determine whether the measurement corresponds to the limit values according to standard EN 12453 and ASR A1.7 or the specifications DHF TS 011 and DHF TS 012 of a DHF measure- ment.
Page 41
► Open the „Details“ tab in the „Measurement“ or „Measurement Wi- zard“ window. ► The calibration date stored in the device can be updated via the „Re- trieving“ button. L Updating the calibration date is required after each new calibration of the KMG-Lite-BT.
Page 42
 Edit customer information of a measurement ► Open the „Details“ tab in the „Measurement“ or „Measurement Wi- zard“ window. ► Use the „Contact“ button to select the desired customer from the list of created customers (see also „Customer managemant/Create customer“, page 31).
Page 43
types and select the type suitable for the measurement. ► Confirm the information with the „Save“ button. Â Set the position of the measurement ► Open the „Details“ tab in the „Measurement“ or „Measurement Wi- zard“ window. ► Select the „Use current position“ button. L To use, your mobile device‘s GPS function must be enabled.
Page 44
 View saved measurements At the start of the KMG-Lite app, the measurement history is available via the „History“ button. In use, you can reach the measurement history via the „Menu field“. Saved measurements are displayed in calendar terms.
Page 45
► Tap the word „Today“ to see one of the possible filters: Today, Yester- day, Last 7 Days, Last 30 Days. ► The display of the data can be used according to „Single“ for single measurements and „Assistent“ for standard measurements or DHF mea- surements.
Page 46
 Create a protocol ► Open the „Protocol“ tab in the „Measurement“ or „Measurement Wi- zard“ window. ► Use the switches to select which information the protocol. L Selecting the „Position“ option requires activating the GPS function of your mobile device. You can display the position as a graphic or as a QR code in the protocol.
Page 47
Error message  Fix error massage „No device found“. ► Reconnect your KMG-Lite-BT to the app. ► If a new error message appears, con- nect your KMG-Lite-BT as a new de- vice.
Care and maintenance L Proper maintenance and care of the equipment maintains its functiona- lity and contributes to the extended life of the product. Â Store device correctly: ► When not used, keep the gauge inside its transport case and protect it from humidity and mechanical stress.
Page 49
L A soiled instrument can cause reduced measurement accuracy. ► In case of doubt send the device to GTE for calibration. As part of the calibration sensitive parts can be cleaned from dust particles. Â Clean housing: ► Clean the housing with a moist cloth (soap suds) if it is dirty. Do not use aggressive cleaning agents or solvents! Â...
Page 50
► Remove old batteries and take care of proper disposal. ► Insert three AA - type batteries with 1.5 V each. ► Mind the polarity! Positive terminals ahead! ► Close battery plug: Press the batteries with the battery plug against the spring resistance inside the handle and lock it by twisting it clockwise.
L To ensure the measurement accuracy the instrument has to be cali- brated every year! L Regular calibration is also mandatory according to DIN 16005 as well as EN 12453. L The calibration badge on the bottom of the device shows when your KMG-Lite- needs to be calibrated.
Page 52
We calibrate your instrument carefully and accurately. Of course, you will receive a factory calibration certificate for each calibration. An order form is included in the scope of delivery of your KMG-Lite. To place an order, send us the completed order form together with your device. GTE...
11 Accessories As additional accessory for KMG-Lite we offer a Spacer-Set, which contains a robust measurement tripod and distance pieces for typical measurement points. Distance elements for Tripod for vertical horizontal gates and doors Opening doors: 30 and 50 cm...
12 Technical data Voltage supply: 3 x 1,5 V - AA - Battery (Mignon) Power consumption: < 25 mA Battery duty: > 100 hours Temperature range: 0 ... 40 °C Humidity range: 20 ... 90 % r. H. (non-condensing) Dimensions of measurement plate: 80 mm Ø, height 50 mm Dimensions: 250 mm x 80 mm x 50 mm (L x W x H)