Instruction Flowmeter Tube - Stephan Artec Service Instructions Manual

Anaesthesia apparatus
Table of Contents

Advertisement

11
List of spare parts
11.9

Instruction Flowmeter tube

Attention
48
Fig. 24: spare parts flowmeter tube
After removing the packing sets (2) and (7),the end plugs (3) and (6) can
be pulled out of the flowmeter tube with the help of a pair of pincers and
the float can be taken out.
The floats, together with the corresponding flowmeter tubes, form a
calibrated system, so that an interchanging must be avoided by all means.
Now the flowmeter tube can be rinsed first with soap suds and then with
clear water.The completely dry tube can be re-installed. After the
plexiglass hood has also been cleaned, the cover of the flowmeter tube
can be screwed up again.
Item
8
Filter for flowmeter tube
Tab. 16: spare parts flowmeter tube
Item 6 must be exchanged on occasion of the semi-annual servicing
SA-103-0109V1.3-HAO-GB
Designation
Article No.
1 924 60 003
© F. Stephan GmbH

Advertisement

Table of Contents
loading

This manual is also suitable for:

Portec

Table of Contents