Flush; Air / N 2 O-Change-Over - Stephan Artec Service Instructions Manual

Anaesthesia apparatus
Table of Contents

Advertisement

3
Description of Design and Performance
3.1.1
O

-Flush

2
3.1.2
AIR / N
12
The high precision flowmeter tube (at the left) indicates the measuring
range from 0 to 2 l/min, the "rough" flowmeter tube (at the right)
indicates the flow quantities from 2 to 15 l/min.
Reading line is the upper edge of the float. Graduation of the lower scale
parts of the measuring tubes (AIR/N
Because the dimensions of the lateral parts of the GME 2/3 can be altered
optionally, the use of flowmeter tubes of other manufacturers (e.g. Rota,
KDG and others) is possible.
Fig. 2: Gas Mixing Unit
Depressing the O
-flush-button effects a quick oxygen supply
2
(approximately 50 l/min) directly to the outlet for fresh gas ( not via the
vaporizer for anaesthetic agent).
Releasing the O
-flush-button effects return to the initial position.
2
O-Change-Over
2
The AIR/N
O-change-over-switch is situated below the respective control
2
knobs (AIR/N
O). It makes preselection of the gases AIR or N
2
possible. The appropriate proportioning of the gases is carried out by
means of the regulating valves O
In case of a decrease of pressure in the supply system (oxygen
lower than 2,8 bar ) an audible alarm sounds for at least 7 seconds.
No muting is possible.
SA-103-0109V1.3-HAO-GB
O) is more closely stepped.
2
-Failure Alarm.
2
© F. Stephan GmbH
O
2