Performance Test; Execution Of The Test - Stephan Artec Service Instructions Manual

Anaesthesia apparatus
Table of Contents

Advertisement

®
7

Performance Test

7.1

Execution of the Test

© F. Stephan GmbH
Fig. 3: Gas Mixing Unit
Proportioning valves (1) on flowmeter unit closed
Cautiously open the proportioning spindle for oxygen while
observing the respiratory pressure gauge (2). Fill pipe line system,
until the respiratory pressure gauge comes to a standstill at constantly
60 mbar.
The float of the oxygen flowmeter now indicates the dimension of the
leakage.
Fig. 4: Proportioning spindle for oxygen
SA-103-0109V1.3-HAO-GB
7
Performance Test