7.7

Test of inspiration-and expiration valve

7.7.1

Basic setting

7.7.2

Execution of the Test

© F. Stephan GmbH
Fig. 8: Test of inspiration-and expiration valve
Circle system in operating mode
Remove mask from Y – piece
Connect the Y - piece with a test lung
Adjust fresh gas flow to 2 l/min
Limit pressure regulating valve to 35 mbar
Fig. 9: Test of inspiration- and expiration valve