7.7
Test of inspiration-and expiration valve
7.7.1
Basic setting
7.7.2
Execution of the Test
© F. Stephan GmbH
Fig. 8: Test of inspiration-and expiration valve
Circle system in operating mode
Remove mask from Y – piece
Connect the Y - piece with a test lung
Adjust fresh gas flow to 2 l/min
Limit pressure regulating valve to 35 mbar
Fig. 9: Test of inspiration- and expiration valve
SA-103-0109V1.3-HAO-GB
7
Performance Test
29