Board D2; Line Filter; Power Pack M 14; Tv Power Supply M13 - Siemens SIREMOBIL Compact System Manual

Table of Contents

Advertisement

Removing the lifting column

Board D2

Remove the protective conductor from the interface plate.
Remove the interface plate.
Remove the grounding bracket from the heat sink. The heat sink remains on the board
(1/Fig.3).
Disconnect all plugs from the board.
Remove the 6 spacer screws.
Unscrew the board from the heat sink (1/Fig.4).
Remove the board and place it on an ESD surface.
Remove the grounding strap from the heat sink (5/Fig.5).
Unscrew the protective conductor.
Remove the 6 mounting screws of the heat sink.
Lift out the heatsink.

Line filter

Disconnect the lower connections of the line filter.
Remove the line filter by removing the screws.
Remove all protective conductors from the base plate of the line filter (2/Fig.5).
Remove the BNC connector (X1) and remove the screws from the BNC plug (3/Fig.5).
Unscrew the grounding strap from the base plate (4/Fig.5).
Remove the screws from the base plate and remove the base plate; remove the bolt from
the lifting column.

Power pack M 14

Disconnect plugs X2 and X3 from the power pack; the protective conductor can remain
connected.
Remove the power pack;
to do this, loosen only the two upper nuts and remove the lower nut.

TV power supply M13

Remove the Plexiglas cover by removing the screws.
Desolder the connections from the transformer and disconnect all other connections.
Remove the screws from underneath the TV power supply and lift it out, Fig.6
Remove the stabilization bracket of the monitor cable connector from the lifting column
by removing the screws , 1/Fig.7.
Siemens AG
Medical Engineering
Register 3
SPR2-130.033.02
Rev. 02
Page 3 of 6
12.98
TD SD 24
2 - 3
2
2
2
2
SIREMOBIL Compact
System Manual