Installing the cover with the control panel
•
Place the cover over the lifting column.
•
Route the ribbon cable and the protective conductor for the control panel through the
opening (1/Fig.9).
•
Put the cover on.
•
Secure the cover to the lifting column (10) with 4 screws and from below the control lever
side with 4 screws as well.
Two long screws are provided on the inside.
•
Reinsert the pin in the control lever.
•
Reinstall the strain relief (2/Fig.9) of the control panel cable.
•
Reattach (3/Fig.9) the protective conductor to the chassis (on the control side from below
left).
•
Reinstall the cover plate (on the control panel side at the chassis, under the control lever)
SIREMOBIL Compact