3 - 6
Fig. 13
1
Fig. 14

Installing the line filter

Insert the 4 standoffs (1/Fig.13) with the short threads into the holes in the lifting column
and tighten them.
Insert the base plate and tighten it with 4 M4 nuts and contact washers.
Pull the cable back behind the lifting column toward the right.
Secure all protective conductors (2/Fig.14) again on the base plate of the line filter;
when doing this, insert contact washers between the cable terminal and the base plate.
Secure the line filter with two screws to the base plate.
Secure the BNC plug to the base plate with the screws.
Connect the BNC cable.

Installing the power pack M 14

Install the base plate.
Install the power pack with the plug connections facing down.
SIREMOBIL Compact
System Manual
2
Register 3
SPR2-130.033.02
Rev. 02
Installing the lifting column