3 - 6
Fig. 13
1
Fig. 14
Installing the line filter
•
Insert the 4 standoffs (1/Fig.13) with the short threads into the holes in the lifting column
and tighten them.
•
Insert the base plate and tighten it with 4 M4 nuts and contact washers.
•
Pull the cable back behind the lifting column toward the right.
•
Secure all protective conductors (2/Fig.14) again on the base plate of the line filter;
when doing this, insert contact washers between the cable terminal and the base plate.
•
Secure the line filter with two screws to the base plate.
•
Secure the BNC plug to the base plate with the screws.
•
Connect the BNC cable.
Installing the power pack M 14
•
Install the base plate.
•
Install the power pack with the plug connections facing down.
SIREMOBIL Compact
System Manual
2
Register 3
SPR2-130.033.02
Rev. 02
Installing the lifting column
Fig. 15
Page 6 of 10