Installing the lifting column
Adjusting the swivel brake
Set point:
•
Move the C-arm into the horizontal position (Fig.19).
•
Measure the braking force with a (20N) spring scale for the released brake and a (200N)
spring scale for the applied brake from the handle grip end to the image intensifier.
(Fig.19);
When doing this, suspend the spring scale with a cloth band or other similar device from
the handle. Set the brake to the above values by adjusting the grip latch on the brake
lever .
•
Place the brake lever in the end positions (brake applied or released) respectively,
parallel to the horizontal arm.
•
Set the brake.
•
Remove the brake lever.
•
Glue the plastic guard in the center with a little silicon adhesive.
•
Install the brake lever in the same position as previously and secure it with the screws.
Remaining work
•
Attach the new ring cover with adhesive to the control lever at the center.
•
Switch the system on and check all functions.
•
Move the lifting column all the way up and attach the scale label supplied vertically in the
middle of the inner section.
•
Install all the covers.
When doing this, attach the protective conductors to the covers.
Perform the protective conductor test.