Installing the lifting column
Plug in the connectors.
Attach the protective conductor to the base plate of the line filter with the screw; insert a
contact washer under the cable terminal.

Installing the TV power supply M13

Insert the TV power supply into the holes in the floor of the unit ; however, don't tighten it
yet.
Pull the cables under the TV power supply and plug in the connections to the board:
Solder the wires to the transformer:
Green/yellow = protective
Lift out the TV power supply once again and install the Plexiglas cover.
Reinstall the TV power supply and secure it from underneath the unit with 4-M4 nuts and
washers.
Secure the protective conductor with the screw to the floor of the unit and to the chassis.
Insert contact washers under the cable terminals.

Installing boards D1 and D2

Secure the heatsink on the control side of the lifting column (Fig.15).
Attach the 6 spacer screws to the heatsink and tighten them (Fig.15).
Apply conductive paste to the contact surface of the heatsink.
Position board D2 on the spacer screws and tighten it to the heatsink (Fig.15).
Connect the plugs on board D2.
Secure board D2 with the spacers Fig.15.
Install the interface plate.
Position the spacers on the 6 screws.
Siemens AG
Medical Engineering
Black wire to 2
conductor
Register 3
SPR2-130.033.02
Rev. 02
Red wire to +
red wire to 0 V
Page 7 of 10
12.98
TD SD 24
yellow wire to 220 V
SIREMOBIL Compact