Installing the lifting column
•
Plug in the connectors.
•
Attach the protective conductor to the base plate of the line filter with the screw; insert a
contact washer under the cable terminal.
Installing the TV power supply M13
•
Insert the TV power supply into the holes in the floor of the unit ; however, don't tighten it
yet.
•
Pull the cables under the TV power supply and plug in the connections to the board:
•
Solder the wires to the transformer:
Green/yellow = protective
•
Lift out the TV power supply once again and install the Plexiglas cover.
•
Reinstall the TV power supply and secure it from underneath the unit with 4-M4 nuts and
washers.
•
Secure the protective conductor with the screw to the floor of the unit and to the chassis.
Insert contact washers under the cable terminals.
Installing boards D1 and D2
•
Secure the heatsink on the control side of the lifting column (Fig.15).
•
Attach the 6 spacer screws to the heatsink and tighten them (Fig.15).
•
Apply conductive paste to the contact surface of the heatsink.
•
Position board D2 on the spacer screws and tighten it to the heatsink (Fig.15).
•
Connect the plugs on board D2.
•
Secure board D2 with the spacers Fig.15.
•
Install the interface plate.
•
Position the spacers on the 6 screws.