Safety Information; Protective Conductor Resistance Test; Other Information; Required Documents - Siemens SIREMOBIL Compact System Manual

Table of Contents

Advertisement

General

Safety information

CAUTION
Tests and adjustments performed with radiation on are identified with the radiation warn-
ing symbol

Protective conductor resistance test

After concluding all work and reattaching all covers, perform the protective conductor test
according to TI 236 / ARTD-002.731.17... .
The protective conductor resistance may not exceed 0.2 ohm.

Other information

n.a.

Required documents

Circuit diagram G5429
Service instructions RXR2-130.061.01....

Required tools and measurement devices

Service kit (refer to ARTD / Part 3)
Digital multimeter
20N spring scale
200N spring scale
25 - 130 Nm 1/2" torque wrench
Service PC
Service-PC - Host connection cable
Protective conductor meter

Required material

Viscogen KL300, 40 g
Conductive paste for heat sink
Lifting column parts kit
Siemens AG
Medical Engineering
When performing service work and tests, adhere to the product-specific
safety information contained in the documents as well as the general
safety information in Register 2 of the TI folder.
. Radiation protection must be worn during these types of adjustments.
Register 3
SPR2-130.033.02
Rev. 02
44 29 114 NH029
44 15 113 RH090
34 24 561
99 00 440 RE999
44 15 899 RV090
73 95 353
37 79 407 G5429
Page 1 of 2
12.98
TD SD 24
1 - 1
1
1
1
1
1
1
1
SIREMOBIL Compact
System Manual

Advertisement

Table of Contents
loading

Table of Contents