Procedure

Dismounting

3
Fig. 1 Dismounting of stereo tactic device
1.
2.
NOTICE
3.
4.
Siemens-Elema AB
Solna, Sweden
X
Y
Z
Remove the four screws holding the front edge of the object table top.
Carefully lift the front edge of the object table top to reveal the four screw
placed behind it.
Do not use excessive force or the object table top
might be damaged.
Remove the revealed screws, the two nuts and the two guiding pins.
Remove the bottom plate by unscrewing the eight remaining screws:
- Two screws at either side of the biopsy unit.
- Four screws at the bottom of the biopsy unit.
Place the biopsy unit as shown in figure 1 (pos. 4)
Register 5
X
Y
Z
SPB7-230.091.09
Rev. 02
01.99
2
4
Object table top
Guiding
pins
Page 1 of 4
MAMMOMAT 3000
SPS-UD
2 - 1
2
1
2
Nuts