Page 3
RIGHTEST GT200 Blood Glucose Monitoring System. This manual provides RIGHTEST GT200 Blood Glucose Monitoring System is a personal blood glucose monitoring system all the information you need to operate this product for accurate test results. Please read this entire to be used for self-testing only. It is not recommended for multiple users.
Page 4
Test Strips. Other brands’ test strips should not be used under any circumstances. The use of based on laboratory test. other brands’ strips may give inaccurate results. - The minimum blood sample size to test using RIGHTEST Blood Glucose Monitoring System is RIGHTEST Blood Glucose Monitoring System is intended for in vitro diagnostic use only.
Page 5
Table of Contents Table of Contents About RIGHTEST Blood Glucose Monitoring System Quality Control About Quality Control Testing......................RIGHTEST Blood Glucose Monitoring System................... Performing a Quality Control Test....................... RIGHTEST Blood Glucose Meter......................Understanding Control Test Results....................RIGHTEST Blood Glucose Test Strip....................
Page 6
RIGHTEST Blood Glucose Monitoring System RIGHTEST Blood Glucose Monitoring System Contents 3* Your RIGHTEST Blood Glucose Monitoring System consists of several items. Please identify each item, learn its name and how it is used. * Below are the items included in your...
Page 7
RIGHTEST Blood Glucose Meter RIGHTEST Blood Glucose Meter Test Strip Port Meter Label Display Window Main Button ( Meter Battery Battery Cover Backlight Battery Right Button ( Left Button (...
Page 8
RIGHTEST Blood Glucose Meter RIGHTEST Blood Glucose Meter Indicates a test result stored in memory Indicates when to apply the blood sample Indicates a test result not included into the test strip average results of the meter Appears when an error occurs...
Page 9
Hold here to Electrochemical sensor PRECAUTION insert test strip - Close the RIGHTEST Blood Glucose Test Strip vial immediately after removing a test strip. into meter. - Do not reuse RIGHTEST Blood Glucose Test Strips. View Window - Do not use expired RIGHTEST Blood Glucose Test Strips.
Page 10
Meter Activation and Battery Change Meter Activation and Battery Change Your RIGHTEST Blood Glucose Meter comes with 2 CR2032, 3 volt, batteries installed. New batteries will provide power to perform approximately 1000 tests under normal use. Press the main button or insert a strip to activate your meter.
Page 11
Press the main button to exit the self-test and enter the Setting Mode. 2. With Battery Inserted Press the main button to turn on RIGHTEST Blood Glucose Meter. Hold down the main button for 2. Month setting 7 seconds.
Page 12
Press the Main button to - If you do no change any settings during Meter Set-up for over 2 minutes, confirm. Once the minute is confirmed, the Volume RIGHTEST Blood Glucose Meter will leave setting mode and power off Setting will appear. automatically.
Page 13
2. Manual Power off 2. Manually turn off the backlight To power off RIGHTEST Blood Glucose Meter, press and hold the Main button for 4 seconds. With the backlight on, press and hold the Main button for another 2 seconds.
Page 14
2. Put your forefinger on the side of the strip as shown. up simultaneously. 3. Insert RIGHTEST Blood Glucose Test Strip into test strip port until it clicks and 3. Take the RIGHTEST Blood Glucose Test Strip out of the test strip port.
Page 15
- RIGHTEST Blood Glucose Meter 2. Pull the cap and hub off in opposite directions, remove - RIGHTEST Blood Glucose Test Strips the cap. (Please check the expiration date on the test strip vial.
Page 16
Close the vial cap immediately. 14. Touch and hold the blood drop to the edge of sample 10. Insert the strip into the test strip port of the RIGHTEST port until the view window is filled with blood. If the view Blood Glucose Meter with the view window facing up.
Page 17
Performing a Blood Glucose Test Performing a Blood Glucose Test 16. Pull off the depth adjustable cap. Without touching the CAUTION used disposable lancet, insert the lancet tip into the - Do not apply your blood drop to the sample port on the strip until you see the " "...
Page 18
RIGHTEST lancing device with Clear Cap for testing sites other than fingertip. - Periodically compare the test system to a laboratory test system that is known to be well - Press the release button.
Page 19
Blood Glucose Test Strip. If this occurs, repeat the test with a new test strip. You can also run a Quality Control Test with the RIGHTEST Control Solutions GC700 to check your Insufficient blood sample Sufficient blood sample RIGHTEST GT200...
Page 20
Monitoring System in the Control Solution Mode. If the test result is within the Control Solution screen. Please repeat your test with by a new test strip. Range printed on the strip vial label, RIGHTEST Blood Glucose Monitoring System passes the Quality Control Test and your RIGHTEST Blood Glucose Monitoring System is working properly.
Page 21
- To prepare for your initial blood glucose test. - To check the RIGHTEST Blood Glucose Test Strip when you open a new vial of strips. Example - To check your RIGHTEST Blood Glucose Meter after it has been dropped, damaged or exposed to liquids.
Page 22
" icon and " " icon on the screen, strip. RIGHTEST Blood Glucose Meter is performing an internal check. Touching the control prompting you to apply the corresponding level of RIGHTEST Control solution to the sample port before prompted will result in an error message: Solution.
Page 23
System may not be working properly. Repeat the Quality Control Test. If your control solution results Control Solution. The control solution test result will not be used when are still out of range, do not use RIGHTEST Blood Glucose Meter to test your blood glucose. Please recalling test average results.
Page 24
Recalling Average Test Results Recalling Average Test Results RIGHTEST Blood Glucose meter provides several average test results. View the 1-day, 7-days, Cancel "exclusion": Stay in the display of "NO AVG". Press the Right or Left button until the 14-days, 30-days, 60-days and 90-days average test results for better blood glucose symbol "NO AVG"...
Page 25
Do not use anything wet to clean it. Perform a quality control test to ensure RIGHTEST Blood Glucose Meter is working properly. The following disinfecting wipe has been tested and may be used to clean and disinfect the meter and lancing device.
Page 26
NOTE NOTE - Your RIGHTEST Lancing Device has been tested to ensure that there is no change in the - Your RIGHTEST Blood Glucose Meter has been tested to ensure that there is no change performance of the device after 550 cleaning cycles and 550 disinfecting cycles. The...
Page 27
- Thin, sliver streaks appear on the screen, 3. If the RIGHTEST Blood Glucose Meter and Test Strips are exposed to a - The screen becomes cracked, soft, dissolved, brittle or swollen.
Page 28
Meter Malfunction again, you may be using the wrong test strip. Please check to make sure you are If the RIGHTEST Blood Glucose Meter will not turn on, please follow the steps below: using a RIGHTEST Test Strip. 1. Open the battery cover and remove the batteries.
Page 29
Limitations Specification - RIGHTEST Blood Glucose Monitoring System is not intended for serum or plasma testing. Measurement Technology Dehydrogenase Electrochemical Sensor - Inaccurate test results may be obtained at altitudes greater than 3,048 meters(10,000 feet) above Measuring Range sea level.
Page 30
65 ± 5g with batteries Meter Weight This warranty does not apply to the performance of a RIGHTEST Blood Glucose Meter that has been altered, misused, tampered with or abused in any way. Monitor This warranty applies only to the original purchaser of the RIGHTEST Blood Glucose Monitoring System.
Page 31
Expected values instructions to make sure you are performing the steps correctly. If you have any questions or in Fasting Blood Glucose case of problems with the RIGHTEST products, please contact local Bionime distributor or email to GLUCOSE LEVEL INDICATION rightest@bionime.com .
Page 32
Component Manufacturer Information Blood Glucose Meter, Test Strip, Control Solution Manufacturer: Bionime Corporation No. 100, Sec.2, Daqing St., South Dist., Taichung City 40242, Taiwan Product complies with In Vitro Diagnostic Medical Device Directive 98/79/EC. ( CE0197 ) EC-Rep: Bionime GmbH, Tramstrasse 16, 9442 Berneck , Switzerland Lancing Device Manufacturer: Bionime Corporation No.
Need help?
Do you have a question about the GT200 and is the answer not in the manual?
Questions and answers