SET USER-DEFINED DEFAULTS
1. Turn unit on
– Review factory settings below
2. Step through ALL settings using
LIMITS button
3. Change desired settings using
or
buttons
4. MUST push
LIMITS button after each
setting to confirm & move on
5. Scroll to end of settings to 2nd
symbol
(
=minutes) and press
LIMITS button
6. Unit will automatically review new settings
7. TO SAVE SETTINGS – press and hold
ALARM SILENCE button while briefly pressing
LIMITS button – "dEF On" will flash
Return to Factory Defaults
Note:
User-Defined Default values are lost
when returned to Factory Defaults.
1. Turn unit on
– "dEF On" will appear
2. Press and hold
ALARM SILENCE button
and briefly press
button – "dEF OFF"
will flash
3. Press
POWER button until unit turns OFF,
then turn back ON
Factory Default Settings
The values at right come pre-programmed in the Model 7500. The
Patient Security Mode is OFF, unless turned on.
Factory
Adjustment
KEY
Parameter
Setting
Options
Increment
SpO
Low Alarm Limit
85%
Off, 50-95
1%
2
Pulse Rate High Alarm Limit
200 BPM
Off, 75-275
5 BPM
Pulse Rate Low Alarm Limit
50 BPM
Off, 30-110
5 BPM
SpO
High Alarm Limit
Off
Off, 80-100
1%
2
Alarm Volume
High
Off, Low, High
N/A
Oximetry Averaging
No
Yes, No
N/A
Model 7500 Quick Start-up Guide
SET PATIENT SECURITY MODE ON
1. Turn unit on
– Review factory settings below
2. Step through ALL settings using
LIMITS button
3. Change desired settings using
or
buttons
4. MUST push
LIMITS button after
each setting
5. Scroll to end of settings to 2nd
symbol
(
=minutes) and press
LIMITS button
6. Unit will automatically review new settings
7. TO SAVE SETTINGS – press and hold
ALARM SILENCE button while briefly pressing
LIMITS button – "dEF On" will flash
8. Press
POWER button until unit turns OFF
9. TO SET – Press and hold
ALARM SILENCE
button while turning the unit back ON
10.Screen will flash "SEC On"
Note:
To set "Low O2" alarm below 85%,
must set in User-Defined Defaults first,
then follow steps above.
Turn Off Patient Security Mode
1. Press
POWER button until unit turns OFF
2. Together hold down
ALARM SILENCE
button, then
LIMITS button and then
POWER button
3. Units reads "SEC OFF"
DATA COLLECTION & DOWNLOAD
Data Collection
1. Turn unit on
– Recording begins
automatically
Note:
Data less than 1 minute is not saved.
Downloading Data To PC
Note:
Patient Security Mode must be OFF
1. Press
POWER button until unit turns OFF
2. Connect unit to PC using 7500SC
download cable
3. Boot up nVISION and select FILE>New Data
Capture, click 7500 oximeter option
4. On 7500 unit press and hold
and
POWER buttons - unit will display
"PLA bAC" during download
5. Unit returns to normal operating mode
after download
Alarm
Power
Silence
(On/Standby)
Sensor
Connector
TO CLEAR MEMORY
1. Turn unit on
Note:
Patient Security Mode must be OFF
2. Press
LIMITS button until "CLr no"
is displayed
3. Press
button until "CLr YES" appears
4. Press
LIMITS button to confirm
5. "dEL no" appears, press
button so
"dEL YES" appears
6. Press
LIMITS button to confirm
7. "dnE CLr" confirms memory has been cleared
For complete specifications, please
refer to the Operator's Manual.
Alarm
%SpO
Pulse Rate
Bar
2
Display
Display
Speaker
Plus (+)
Limits
Minus (-)
(step through settings)