E. SPAN Width
The value of this is determined by the SPAN WIDTH Knob and shows the MHz/div chosen.
F. MENU SELECTION Value
This shows the value of the chosen menu selection. In the above reference, it shows the value of
the screen contrast. As the CONTRAST + menu button is pressed, the value will increase.
G. AMPLITUDE
The numbers on the right side of the screen depict the amplitude value in dB.
H. MINIMUM FREQUENCY Value
A value in this position shows the lower frequency value from the center line. It is derived from the
Center Frequency — (Span Width/div X 4 Div)
I. TIME
This is the current time.
J. CENTER Frequency
Center Frequency as determined by the TUNING Knob.
K. DATE
This is the current date as set in the analyzer.
L. MAXIMUM Frequency
A value in this position shows the upper frequency value from the center line. It is derived from the
Center Frequency + (Span Width/div X 4 Div)